The constrained version of generalized predictive control (GPC) which employs the quadratic programming (QP) approach is evaluated for on-line administration of an anaesthetic drug in the operating theatre during surgery. In the first instance, a patient simulator was developed using a physiological model of the patient and the necessary control software was validated via a series of extensive simulation experiments. Such a validated system was then transferred into the operating theatre for a series of clinical evaluation trials. The clinical trials, which were performed with little involvement of the design engineer, led to a good regulation of unconsciousness using fixed-parameters as well the adaptive version of the algorithm. Furthermore, the constrained algorithm displayed good robustness properties against disturbances such as high stimulus levels and allowed for safe and economically effective administration of the anaesthetic agent isoflurane.
机译:评估了采用二次规划(QP)方法的广义预测控制(GPC)的受限版本,以在手术过程中对手术室中的麻醉药物进行在线给药。首先,使用患者的生理模型开发了患者模拟器,并通过一系列广泛的模拟实验验证了必要的控制软件。然后将经过验证的系统转移到手术室中进行一系列临床评估试验。在设计工程师的很少参与下进行的临床试验,通过使用固定参数以及算法的自适应版本,可以很好地调节意识。此外,受约束的算法对干扰(例如高刺激水平)显示出良好的鲁棒性,并允许安全,经济地管理麻醉剂异氟烷。