首页> 外文期刊>Instrumentation and Measurement, IEEE Transactions on >On the Calibration of Direct-Current Current Transformers (DCCT)
【24h】

On the Calibration of Direct-Current Current Transformers (DCCT)

机译:关于直流电流互感器(DCCT)的校准

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Modern commercial dc current transformers (DCCTs) can measure currents up to the kiloampere range with accuracies better than . We discuss here a DCCT calibration method and its implementation with commercial instruments typically employed in low-resistance calibration laboratories. The primary current ranges up to 2 kA; in the current range <100 A, the calibration uncertainty is better than . An example of calibration of a high-performance DCCT specified for primary currents measurement up to 900 A is discussed in detail.
机译:现代商用直流电流互感器(DCCT)可以测量高达千安培范围的电流,其精度优于。我们在这里讨论DCCT校准方法及其在低电阻校准实验室中通常使用的商用仪器中的实现。初级电流范围高达2 kA;在<100 A的电流范围内,校准不确定度优于。详细讨论了为高达900 A的一次电流测量而指定的高性能DCCT的校准示例。