In September 1984, the 54 saturated standard cells which had been used to maintain the Fluke corporate volt were removed from service and placed in storage. Since that time, Fluke''s corporate volt has been maintained in 4 solid-state, 10-V references. Discarding the cells has much reduced the time and cost of maintaining the standard, has significantly decreased its uncertainty, and has much improved the confidence that stated uncertainties are actually attained. An uncertainty of +/− 0.08 ppm was assigned to the difference between 10 V at Fluke and 10 V as maintained by NBS and an uncertainty of ±0.40 ppm to the absolute value of the Fluke 10 V. This paper describes the hardware and the statistics utilized in establishing the solid state primary standard of voltage and reports the results attained to date.
机译:1984年9月,将用于维持Fluke公司电压的54个饱和标准电池从服务中移除,并进行了存储。从那时起,Fluke的公司电压一直保持在4个固态10V参考电压中。丢弃细胞大大减少了维持标准液的时间和成本,显着降低了其不确定性,并且大大提高了确定的不确定性实际上得到的可信度。在Fluke时10 V和由NBS维持的10 V之间的差值中,+ /-0.08 ppm的不确定度和Fluke 10 V的绝对值的±0.40 ppm不确定度。本文描述了硬件和统计数据用于建立电压的固态主要标准并报告迄今获得的结果。