Development of a Simple In Vitro Artery Model and an Evaluation of the Impact of Pulsed Flow on High-Intensity Focused Ultrasound Ablation

机译:一种简单的体外动脉模型的开发和对高强度聚焦超声消融脉冲流动影响的评价

获取原文
获取原文并翻译 | 示例

摘要

Background: Many parameters have to be investigated before an optimal strategy for thermal therapies can be defined. Such studies are limited by the number of animals that may be included and by the difficulties in the use of isolated perfused organs. Realistic in vitro anthropomorphic vascular phantoms have been suggested and are based on the use of agar or silicone but not on biological tissues. More simply, biological and reliable models can be developed to mimic the behavior of arteries under perfusion.Material and methods: Chicken esophagus was used since it reproduces both the anatomical and the mechanical local properties of a 5-mm-diameter artery. The esophagus was placed inside a bovine liver and connected to polyurethane tubes. The flow was driven using two pumps and a solenoid to mimic pulsed flow and the cardiac valve. The blood was mimicked using degassed water and 50-mu m silica beads at a concentration of 40 mg/L. The Doppler signals used as a reference to validate the in vitro artery model were acquired during an animal study. Using a high-intensity focused ultrasound (HIFU) device, several HIFU lesions were created in vitro in liver samples with the artery mimic model placed at different depths.Results: The waveforms of in vivo and in vitro Doppler signals had a cycle length of 1.09 s and 1.10 s, respectively, corresponding to 55 beats per minute (bpm). The average peak flow rate was 25.3 cm/s for in vivo waveforms and 27.8 cm/s for in vitro waveforms. The relative distension of the in vitro artery mimic (10%) was similar to that measured in pig mesenteric arteries and representative of human artery compliance. The dimensions of HIFU lesions were different depending on the location of the artery mimic.Conclusions: A simple and reliable model of arteries is described. This model reproduced both the geometrical and the mechanical local properties of an artery. The flow profile, the flow rate and compliant behavior were found to be similar to those that can be recorded in vivo. This model can be used to evaluate the potential perfusion effects when developing devices for thermal therapies. (C) 2020 AGBM. Published by Elsevier Masson SAS. All rights reserved.
机译:背景:必须在可以定义热疗法的最佳策略之前进行许多参数。这些研究受到可包括的动物的数量,并且通过使用分离的灌注器官使用的困难。已经提出了现实的体外拟人血管血管脉模,并基于琼脂或硅树脂但不在生物组织上的使用。更简单地,可以开发生物和可靠的模型以模仿灌注下的动脉的行为。材料和方法:使用鸡食管,因为它再现了5毫米直径动脉的解剖和机械局部性质。将食道放入牛肝内并连接到聚氨酯管。使用两个泵和螺线管驱动流动以模仿脉冲流量和心脏瓣膜。使用脱气的水和50-mu m二氧化硅珠,以40mg / L的浓度模仿血液。在动物研究期间获得了作为参考验证体外动脉模型的多普勒信号。使用高强度聚焦超声(HIFU)器件,在肝脏样品中在体外产生几个HIFU病变,其中动脉模拟模拟模型放置在不同的深度。结果:体内和体外多普勒信号的波形具有1.09的循环长度为1.09 S和1.10 S分别对应于每分钟55次(BPM)。体内波形的平均峰值流速为25.3cm / s,用于体外波形27.8cm / s。体外动脉模拟(10%)的相对扩散与在猪肠系膜动脉中测量的相对扩张和人类动脉依从性的代表。根据动脉模拟的位置,HIFU病变的尺寸不同。结论:描述了一种简单且可靠的动脉模型。该模型再现动脉的几何和机械局部特性。发现流程分布,流速和柔顺行为类似于可以在体内记录的行为。该模型可用于在开发热疗法的装置时评估潜在的灌注效果。 (c)2020年AGBM。由Elsevier Masson SA出版。版权所有。

著录项

  • 来源
  • 作者单位

    Univ Lyon 1 Univ Lyon INSERM LabTAU Ctr Leon Berard F-69003 Lyon France|EDAP TMS 4 Rue Dauphine F-69120 Vaulx En Velin France;

    Univ Lyon 1 Univ Lyon INSERM LabTAU Ctr Leon Berard F-69003 Lyon France;

    EDAP TMS 4 Rue Dauphine F-69120 Vaulx En Velin France;

    Univ Lyon 1 Univ Lyon INSERM LabTAU Ctr Leon Berard F-69003 Lyon France;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号