A simulation study of sample volume sensitivity for oblique pulsed finite beam insonation of Doppler ultrasound flow phantom cylindrical vessels

机译:多普勒超声流幻影圆柱状血管斜脉冲有限束超声成像的样品体积敏感性模拟研究

获取原文
获取原文并翻译 | 示例
           

摘要

Our previous analysis of the lumen pressure in Doppler ultrasound flow phantoms subject to continuous wave, infinite beam excitation is extended here to consider the pressure and Doppler sample volume complex sensitivity within a range of solid absorbent tubes typical of those used in Doppler ultrasound flow phantoms insonated with a focussed pulsed ultrasound beam. The beam may be incident on the cylindrical shell from any angle and with any offset from the shell axis. The examples considered are of a 5 MHz beam with a 6 dB lateral fullwidth of 1 mm at the focus and a transducer surface acceleration pulse with standard deviation of 1 /spl mu/s propagating through 10 mm outer diameter, 8 mm inner diameter, Cflex, low-density polyethylene (LDPE), high-density polyethylene (HDPE), and poly-methylmethacrylate (PMMA) shells surrounded by water at various beam-vessel angles. Our results confirm earlier analyses suggesting that PMMA, being less well matched to the surrounding media, causes much greater distortion of the sample volume sensitivity than Cflex.
机译:我们先前对连续波多普勒超声流幻影中的管腔压力的分析,这里扩展了无限束激发,以考虑在多普勒超声流幻影中典型使用的固体吸收管范围内的压力和多普勒样品体积的复数灵敏度。聚焦脉冲超声束。光束可以从任何角度以任​​何角度入射到圆柱外壳上,并且与外壳轴线之间的偏移都可以。所考虑的示例是5 MHz光束,焦点处的横向全宽为6 dB,其焦点为1 mm,传播范围为10 mm,内径为8 mm,Cflex的换能器表面加速度脉冲的标准偏差为1 / spl mu / s。 ,低密度聚乙烯(LDPE),高密度聚乙烯(HDPE)和聚甲基丙烯酸甲酯(PMMA)外壳,它们在各种光束角度下都被水包围。我们的结果证实了较早的分析结果,表明与周围介质的匹配较差的PMMA会比Cflex引起更大的样品体积灵敏度失真。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号