首页> 外文期刊>IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control >Real-time rectilinear 3-D ultrasound using receive mode multiplexing
【24h】

Real-time rectilinear 3-D ultrasound using receive mode multiplexing

机译:使用接收模式多路复用的实时直线3-D超声

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

In previous work, we developed two generations of a real-time rectilinear volumetric scanner operating at 5 MHz for abdominal, breast, or vascular imaging using a Mills cross two-dimensional (2-D) array and a rectilinear periodic 2-D array. To improve spatial resolution performance and sensitivity, we developed a new design using 4:1 receive mode multiplexing. With 4:1 multiplexing, the new 65,000 element 2-D array has 4/spl times/256=1024 receivers so that 256 receivers can be used on any image line. The two major benefits of using receive mode multiplexing are an increase in receive sensitivity due to a greater number of receive elements, and a decrease in grating lobe and clutter levels due to increased receive element density. Theoretical simulations and analysis show an increase of about 13 dB in sensitivity compared to our previous work. With these encouraging results, a new 65,000 element 5-MHz, 2-D array having 1024 receivers and 169 transmitters was prototyped. In addition, the multiplexer and control circuitry were designed, built, and interfaced with both the transducer and volumetric scanner. Images of tissue-mimicking phantoms and in vivo targets were obtained. Using a spherical cyst phantom, experimental results showed a+12 dB improvement in signal-to-noise ratio and a+6 dB improvement in contrast compared to our previous work.
机译:在以前的工作中,我们使用Mills交叉二维(2-D)阵列和直线周期性2-D阵列,开发了两代实时直线体积扫描仪,以5 MHz的频率进行腹部,乳房或血管成像。为了提高空间分辨率的性能和灵敏度,我们开发了一种使用4:1接收模式多路复用的新设计。通过4:1复用,新的65,000元素2-D阵列具有4 / spl次/ 256 = 1024个接收器,因此在任何图像线上均可使用256个接收器。使用接收模式多路复用的两个主要好处是:由于接收元件数量增加,接收灵敏度提高;由于接收元件密度增加,光栅波瓣和杂波电平降低。理论仿真和分析表明,与我们以前的工作相比,灵敏度提高了约13 dB。有了这些令人鼓舞的结果,新的具有65,000个元素的5 MHz,二维阵列具有1024个接收器和169个发送器的原型。此外,还设计,构建了多路复用器和控制电路,并将其与换能器和体积扫描仪连接。获得了模仿组织的幻像和体内靶标的图像。与我们以前的工作相比,使用球形囊状体模,实验结果表明信噪比提高了+12 dB,而对比度提高了+6 dB。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号