PETRRA: preliminary experimental results from the first full size detector and dead time simulation of the count rate performance of a unique whole body PET camera

机译:PETRA:首款全尺寸探测器的初步实验结果以及独特的全身PET相机计数率性能的停滞时间模拟

获取原文
获取原文并翻译 | 示例
           

摘要

The design and construction of the first of two BaF/sub 2/-TMAE large area detectors to be incorporated in a unique double headed whole body PET system (PETRRA) has been completed. Preliminary experimental work carried out with the first large area detector shows an absolute detection efficiency of 20.5% and a time resolution of 3.5 ns (FWHM) and 8.0 ns (FWTM), using 8 mm thick scintillation crystals. An analytical model of the electronic read-out system has been developed to investigate the rate capability of the new scanner. This model predicts that under typical imaging conditions, with 40 MBq of activity in the field of view, the proposed scanner will acquire coincidence data at 20-25 kcps.
机译:已完成将两个BaF / sub 2 / -TMAE大面积探测器中的第一个结合到独特的双头全身PET系统(PETRRA)中的设计和构造。使用8 mm厚的闪烁晶体,使用第一个大面积检测器进行的初步实验工作显示绝对检测效率为20.5%,时间分辨率为3.5 ns(FWHM)和8.0 ns(FWTM)。已经开发出电子读出系统的分析模型,以研究新扫描仪的速率能力。该模型预测,在典型的成像条件下,视场活动为40 MBq时,建议的扫描仪将以20-25 kcps的速度获取重合数据。