Basic performance of the PET scanner, EXACT HR/sup +/, with adjustable data-acquisition parameters

机译:PET扫描仪的基本性能,EXACT HR / sup + /,具有可调节的数据采集参数

获取原文
获取原文并翻译 | 示例
           

摘要

The sensitivity and the spatial resolution of the ECAT EXACT HR/sup +/, a 3D PET scanner from Siemens/CTI, were analyzed with reference to adjustable data acquisition parameters. The standard protocol of the National Electrical Manufacturers Association (NEMA) was followed with a minor modification. The performance was shown to depend on the parameters prominently in the 3D mode. The advantages and disadvantages of altering the parameters were discussed. It was concluded that the data acquisition parameters should be carefully selected for a specific clinical situation in order to maximize the efficiency of PET measurement.
机译:参照可调数据采集参数,分析了Siemens / CTI的3D PET扫描仪ECAT EXACT HR / sup + /的灵敏度和空间分辨率。遵循美国国家电气制造商协会(NEMA)的标准协议进行了少量修改。在3D模式下,显示性能主要取决于参数。讨论了更改参数的优缺点。结论是,应针对特定的临床情况仔细选择数据采集参数,以最大程度地提高PET测量的效率。