Evaluation of ZnS:6LiF and ZnO:6LiF Scintillation Neutron Detectors Readout With SiPMs

机译:SiPMs对ZnS: 6 LiF和ZnO: 6 LiF闪烁中子探测器的评估

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

It is generally accepted that the count rate capability of ZnS:n6nLiF neutron detectors is strongly limited due to the long afterglow of the scintillator. However, realized as an array of individual sensitive elements (pixels) with independent signal readout and processing, such detectors can be operated with blocking times down to 1n$mu text{s}$nand promising counting rates up to 100 kHz/pixel. In this paper, we investigate how the long emission of the ZnS scintillator influences the detector performance apart from requiring a finite blocking time to suppress the probability of multiple counts. We also study the potential of the ZnO:n6nLiF scintillator with lower afterglow as a possible replacement of ZnS:n6nLiF.
机译:普遍认为ZnS:n 6 nLiF中子探测器由于闪烁体的长余辉而受到严格限制。但是,由于实现为具有独立信号读取和处理功能的单个敏感元件(像素)的阵列,因此此类检测器可以在低至1n的阻塞时间内运行。 $ mu text {s} $ < / inline-formula> nand有希望的计数率高达100 kHz /像素。在本文中,我们研究了ZnS闪烁体的长发射如何影响检测器性能,除了需要有限的阻塞时间来抑制多次计数的可能性。我们还研究了ZnO:n 6 nLiF闪烁体具有较低的余辉,可以替代ZnS:n 6 nLiF。