首页> 外文期刊>IEEE Transactions on Medical Imaging >Automatic Resonance Frequency Retuning of Stretchable Liquid Metal Receive Coil for Magnetic Resonance Imaging
【24h】

Automatic Resonance Frequency Retuning of Stretchable Liquid Metal Receive Coil for Magnetic Resonance Imaging

机译:用于磁共振成像的可拉伸液态金属接收线圈的自动谐振频率垂直

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Stretchable magnetic resonance (MR) receive coils show shifts in their resonance frequency when stretched. An in-field receiver measures the frequency response of a stretchable coil. The receiver and coil are designed to operate at 128 MHz for a 3T MR scanner. Based on the measured frequency response, we are able to detect the changes of the resonance frequency of the coil. We show a proportional-integral-derivative controller that tracks the changes in resonance frequency and retunes the stretchable coil. The settling time of the control loop is less than 3.8ms. The retuning system reduces the loss in signal-to-noise ratio of phantom images from 1.6 dB to 0.3 dB, when the coil is stretched by 40% and the coil is retuned to 128 MHz.
机译:拉伸磁共振(MR)接收线圈在拉伸时显示它们的共振频率。现场接收器测量可拉伸线圈的频率响应。接收器和线圈设计成在128 MHz上为3T MR扫描仪操作。基于测量的频率响应,我们能够检测线圈的谐振频率的变化。我们示出了比例 - 积分 - 衍生控制器,其跟踪谐振频率的变化并重新旋转可拉伸线圈。控制回路的稳定时间小于3.8ms。当线圈拉伸40%时,重新调度系统减少了从1.6dB的幻影图像的噪声比到0.3dB的损失。