Evaluation of the ECAT EXACT HR+ 3-D PET scanner in H/sub 2//sup 15/O brain activation studies: dose fractionation strategies for rCBF and signal enhancing protocols

机译:在H / sub 2 // sup 15 / O大脑激活研究中评估ECAT EXACT HR + 3-D PET扫描仪:rCBF和信号增强方案的剂量分割策略

获取原文
获取原文并翻译 | 示例

摘要

The authors evaluated the performance of the ECAT EXACT HR+ 3-D whole-body positron emission tomography (PET) scanner when employed to measure brain function using H/sub 2//sup 15/O bolus activation protocols that are completed in single same-day data acquisition sessions. Using vibrotactile and auditory stimuli as independent activation tasks, the authors studied the scanner performance under different imaging conditions in five healthy volunteers. Cerebral blood flow images were acquired from each volunteer using H/sub 2//sup 15/O bolus injections of activity varying from 5-20 mCi. One-session dose-fractionation strategies were analyzed for rCBF, standard activity-concentration, switched, and cold-bolus/switched protocols. Performance characteristics: The scanner dead time grew linearly with injected dose from 10% to 25%. Random events varied from 30% to 50% of the detected events. Random and scattered events were corrected adequately at all doses. Estimated noise effective-count curves plateau at about 10 mCi. One-session 12-injection bolus PET activation protocols. Using an acquisition protocol that accounts for the scanner performance and the practical aspects of imaging volunteers and neurological patients in a single same-day session, the authors assessed the correlation between the significance of activation foci and the dose/injection used. The one session protocol employs 12 bolus injections/subject. The authors present evidence suggesting that when an rCBF protocol is used, image noise is reduced significantly when the activity injected increases from 5 to 10 mCi. Increasing the dose from 10 to 15 or 20 mCi yielded further but smaller reductions. The authors' observations also suggest that image noise will be strongly reduced if a 20-mCi dose/injection is used when data are collected using protocols that employ long acquisition times such as a switched or a cold-bolus/switched protocol.
机译:作者评估了ECAT EXACT HR + 3-D全身正电子发射断层扫描(PET)扫描仪在使用H / sub 2 // sup 15 / O推注激活方案测量脑功能时的性能,该方案在单个相同的条件下完成日数据采集会话。使用震动触觉和听觉刺激作为独立的激活任务,作者研究了五名健康志愿者在不同成像条件下的扫描仪性能。使用H / sub 2 // sup 15 / O大剂量注射从5-20 mCi变化的活性从每位志愿者获取脑血流图像。分析了一疗程剂量分级策略的rCBF,标准活动浓度,转换和冷推注/转换方案。性能特征:扫描仪的死区时间随着注入剂量的10%到25%线性增长。随机事件从检测到的事件的30%到50%不等。在所有剂量下,随机事件和零星事件均得到适当纠正。估计的噪声有效计数曲线稳定在大约10 mCi。一站式12针推注PET激活方案。作者使用了一项采集协议,该协议考虑了扫描仪的性能以及在同一天的会议中对成像志愿者和神经病患者进行成像的实际情况,作者评估了激活灶的重要性与所用剂量/注射剂量之间的相关性。一个会话方案采用12次推注/受试者。作者提供的证据表明,当使用rCBF方案时,当注入的活性从5 mCi增加到10 mCi时,图像噪声会大大降低。将剂量从10 mCi增加到15或20 mCi会产生进一步但较小的减少。作者的观察结果还表明,如果使用采用长采集时间的协议(如交换协议或冷推注/交换协议)收集数据时,如果使用20-mCi剂量/注射,图像噪声将大大降低。

著录项

相似文献

  • 外文文献
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号