To improve the SpO(2) sensing system performance for hypoperfusion (low perfusion index) applications, this paper proposes a low-noise light-to-frequency converter scheme from two aspects. First, a low-noise photocurrent buffer is proposed by reducing the amplifier noise floor with a transconductance-boost (g(m)-boost) circuit structure. Second, a digital processing unit of pulse-frequency-duty-cycle modulation is proposed to minimize the quantization noise in the following timer by limiting the maximum output frequency. The proposed light-to-frequency sensor chip is designed and fabricated with a 0.35-mu m CMOS process. The overall chip area is 1 x 0.9 mm(2) and the typical total current consumption is about 1.8 mA from a 3.3-V power supply at room temperature. The measurement results prove the proposed functionality of output pulse duty cycle modulation, while the SNR of a typical 10-kHz output frequency is 59 dB with about 9-dB improvement when compared with the previous design. Among them, 2-3 dB SNR improvement stems from the gm-boosting and the rest comes from the layout design. In-system experimental results show that the minimum measurable PI using the proposed blood SpO(2) sensor could be as low as 0.06% with 2-percentage-point error of SpO(2). The proposed chip is suitable for portable low-power high-performance blood oximeter devices especially for hypoperfusion applications.
机译:为了提高SPO(2)感测系统性能的低灌注(低灌注指数)应用,本文提出了两个方面的低噪声光到频转换器方案。首先,通过用跨导 - 升压(G(m)-boost)电路结构减少放大器噪声底板来提出低噪声光电流缓冲器。其次,建议通过限制最大输出频率来最小化以下定时器中的量化噪声的数字处理单元。所提出的光到频传感器芯片采用0.35 - MU M CMOS工艺设计和制造。整体芯片面积为1×0.9mm(2),典型的总电流消耗量在室温下的3.3V电源约为1.8 mA。测量结果证明了输出脉冲占空比调制的拟议功能,而典型的10-kHz输出频率的SNR是59dB,与先前的设计相比,大约9-DB的改进。其中,2-3 dB SNR改善源于通用汽车升压,其余部分来自布局设计。在系统中的实验结果表明,使用所提出的血液SPO(2)传感器的最小可测量PI可能低至0.06%,SPO(2)的2百分点误差为0.06%。所提出的芯片适用于便携式低功率高性能血液血氧仪装置,尤其适用于低贮料应用。

著录项

  • 来源
  • 作者单位

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Guangdong Biolight Meditech Co Ltd Zhuhai 519085 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Chongqing Univ Sch Microelect & Commun Engn Chongqing Engn Lab High Performance Integrated Ci Chongqing 400044 Peoples R China;

    Hamad Bin Khalifa Univ Coll Sci & Engn Doha 34110 Qatar;

    Hamad Bin Khalifa Univ Coll Sci & Engn Doha 34110 Qatar;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Blood SpO(2); low noise; hypoperfusion; light-to-frequency converter; portable biomedical sensor;

    机译:血液孢子(2);低噪声;低灌注;光到变频器;便携式生物医学传感器;