摘要

Objective: To close perimembranous ventricular septal defects (PMVSDs) in children with the new Amplatzer asymmetric ventricular septal defect occluder (AAVSDO). Patients and design: 10 children, aged 1.5―12 years, with PMVSDs underwent transcatheter closure with the AAVSDO. The device consists of two low profile disks made of Nitinol wire mesh with a 1.5 mm connecting waist. The left disk is 5 mm towards the apex and only 0.5 mm towards the aortic valve. The right disk is 4 mm larger than the waist. The prosthesis diameter was chosen to be 1―2 mm larger than the largest diameter of the defect (determined by transoesophageal echocardiography and angled angiocardiography). A 7―8 French gauge sheath was used to deliver the AAVSDO. Results: The PMVSD diameter ranged from 2―8 mm. The device diameter ranged from 4―8 mm. After deployment of the prosthesis there was no residual shunt in 9 of 10 patients (90%). In one patient there was a trivial residual shunt that disappeared at the three month follow up. Three patients developed transient complete left bundle branch block. No other complications were observed. Conclusions: The AAVSDO appears to be a promising device for transcatheter closure of PMVSDs in children. Further studies are required to document its efficacy, safety, and long term results in a larger patient population.
机译:目的:采用新型Amplatzer非对称性室间隔缺损封堵器(AAVSDO)封闭儿童的膜周室间隔缺损(PMVSDs)。患者和设计:10例1.5至12岁的PMVSD患儿接受AAVSDO封堵术。该设备包括两个由镍钛诺金属丝网制成的低矮型圆盘,其连接腰部为1.5毫米。左圆盘朝向顶点5毫米,而朝向主动脉瓣仅0.5毫米。右圆盘比腰部大4毫米。假体直径选择为比缺损的最大直径大1–2 mm(由经食道超声心动图和成角度的心动图确定)。使用7-8法式量规护套来发射AAVSDO。结果:PMVSD直径范围为2-8 mm。设备直径范围为4-8 mm。部署假体后,在10名患者中有9名(90%)没有残余分流。在一名患者中,有一个琐碎的残留分流器,在三个月的随访中消失了。三例患者出现暂时性完全性左束支传导阻滞。没有观察到其他并发症。结论:AAVSDO似乎是一种有希望的用于经导管闭合儿童PMVSD的设备。需要进一步的研究来证明其功效,安全性以及在更大的患者人群中的长期结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号