We present the first cadavic study results concerning the feasibility of the use of an MR-guided assistance system, Innomotion (Innomedic, Herxheim, Germany), for accurate and consistent placement of percutaneous needles in the prostate gland. The MR-compatible assistance system consists of a C-arch, guiding arm and application module (AMO). T1-weighted fast low angle shot (FLASH) 2-D-GRE sequence (TR/TE=110/4 ms) and T2-weighted turbo spin-echo (TSE)-sequences (TR/TE=3200/97 ms) in transversal orientation were used for the monitoring of the punction of the prostate gland. Planning and control of the intervention is to be made outside the scanner room on a desktop computer that receives DICOM images from the scanner. Servopneumatic drives move the AMO to the insertion point. The physician has to introduce the punction needle manually. The mean deviation of the needle tip to the target in a gel phantom was 0.35 mm. An accurate punction of the prostate gland can easily be performed using this system with a transgluteal access. The T2-weighted images are superior for the evaluation of the prostate anatomy and the needle position during the interventions. In conclusion, our preliminary results indicate that this MR-guided assistance system is suitable for an accurate transgluteal needle placement in the prostate.
机译:我们提出了有关使用MR引导辅助系统Innomotion(Innoedic,德国赫克斯海姆,德国)以将前列腺穿刺针准确且一致地放置的可行性的第一个cadavic研究结果。与MR兼容的辅助系统由C形弓,导向臂和应用模块(AMO)组成。 T1加权快速低角度拍摄(FLASH)2-D-GRE序列(TR / TE = 110/4 ms)和T2加权涡轮自旋回波(TSE)序列(TR / TE = 3200/97 ms)横向定向用于监测前列腺的点刺。干预措施的计划和控制应在台式机的扫描仪室外进行,该台式计算机从扫描仪接收DICOM图像。伺服气动驱动器将AMO移至插入点。医师必须手动引入穿刺针。在凝胶体模中,针尖到目标的平均偏差为0.35 mm。使用带有经臀的通路的该系统,可以轻松地对前列腺进行精确的点刺。 T2加权图像在干预过程中对前列腺解剖结构和针头位置的评估非常出色。总之,我们的初步结果表明,这种MR引导辅助系统适用于在前列腺中准确地经臀置入针头。

著录项

  • 来源
    《European Radiology》 |2007年第4期|1118-1124|共7页
  • 作者单位

    Institute for Diagnostic and Interventional Radiology University Hospital Frankfurt Johann Wolfgang Goethe-University Theodor-Stern-Kai 7 60590 Frankfurt am Main Germany;

    Institute for Diagnostic and Interventional Radiology University Hospital Frankfurt Johann Wolfgang Goethe-University Theodor-Stern-Kai 7 60590 Frankfurt am Main Germany;

    Institute for Diagnostic and Interventional Radiology University Hospital Frankfurt Johann Wolfgang Goethe-University Theodor-Stern-Kai 7 60590 Frankfurt am Main Germany;

    Institute for Diagnostic and Interventional Radiology University Hospital Frankfurt Johann Wolfgang Goethe-University Theodor-Stern-Kai 7 60590 Frankfurt am Main Germany;

    Innomedic GmbH Luitpoldstraße 59 76863 Herxheim Germany;

    Innomedic GmbH Luitpoldstraße 59 76863 Herxheim Germany;

    Innomedic GmbH Luitpoldstraße 59 76863 Herxheim Germany;

    Department of Radiology Medical University of South Carolina 169 Ashley Avenue Charleston SC 29425 USA;

    Department of Radiology Medical University of South Carolina 169 Ashley Avenue Charleston SC 29425 USA;

    Institute for Diagnostic and Interventional Radiology University Hospital Frankfurt Johann Wolfgang Goethe-University Theodor-Stern-Kai 7 60590 Frankfurt am Main Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Magnetic resonance (MR); Prostate neoplasms; Guidance;