首页> 外文期刊>ELECTROPHORESIS >Microfluidic devices with disposable enzyme electrode for electrochemical monitoring of glucose concentrations†
【24h】

Microfluidic devices with disposable enzyme electrode for electrochemical monitoring of glucose concentrations†

机译:具有一次性酶电极的微流体装置,用于电化学监测葡萄糖浓度†

获取原文
获取原文并翻译 | 示例
           

摘要

This article describes the fabrication of tube-like microchannels made of UV curable polymer on a glass substrate and the device assembling with a disposable enzyme-working electrode for high-sensitivity electrochemical detection. While both reference and counter electrodes are patterned on the surface of the glass substrate, the working electrode is flipped on the top of the channel with an open access, providing a face-to-face probing configuration. When the enzyme electrode is contaminated or degraded, it can be easily replaced by a new one, keeping the main body of the device and the detection schema unchanged. Using glucose oxidase-coated gold electrodes, we were able to determine a linear amperometry response to the glucose concentrations in the range of 2–16 mM. By replacing the as-prepared working electrode by the one after thermal treatments, we showed a much more degraded enzyme electrode activity, enabling efficient determination of the electrode quality as well as the whole process optimization.
机译:本文介绍了在玻璃基板上制造由紫外线固化聚合物制成的管状微通道,以及与用于高灵敏度电化学检测的一次性酶工作电极组装在一起的设备。当参比电极和对电极都在玻璃基板的表面上进行构图时,工作电极在通道的顶部翻转,具有开放通道,从而提供了面对面的探测配置。当酶电极被污染或降解时,可以很容易地用新的电极替换,从而使装置的主体和检测方案保持不变。使用涂有葡萄糖氧化酶的金电极,我们能够确定对2–16µmM范围内的葡萄糖浓度的线性安培法响应。通过用热处理后的电极替换准备好的工作电极,我们发现酶电极的活性大大降低,从而可以有效地确定电极质量并优化整个过程。

著录项

  • 来源
    《ELECTROPHORESIS》 |2011年第22期|p.3201-3206|共6页
  • 作者单位

    Ecole Normale Supérieure, Paris, France|Institute for Integrated Cell-Material Sciences, Kyoto University, Kyoto, Japan|East China Normal University, Shanghai, P. R. China;

    |Ecole Normale Supérieure, Paris, France|East China Normal University, Shanghai, P. R. China;

    |Ecole Normale Supérieure, Paris, France;

    Ecole Normale Supérieure, Paris, France|Institute for Integrated Cell-Material Sciences, Kyoto University, Kyoto, Japan;

    Ecole Normale Supérieure, Paris, France;

    Ecole Normale Supérieure, Paris, France;

    Ecole Normale Supérieure, Paris, France;

    Institute for Integrated Cell-Material Sciences, Kyoto University, Kyoto, Japan;

    East China Normal University, Shanghai, P. R. China;

    East China Normal University, Shanghai, P. R. China;

    Ecole Normale Supérieure, Paris, France|Institute for Integrated Cell-Material Sciences, Kyoto University, Kyoto, Japan;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Electrochemical detection; Microdevices; Microfluidics;

    机译:电化学检测;微器件;微流体;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号