Morning pulmonary artery pressure measurements by CardioMEMS are most stable and recommended for?pressure trends monitoring

机译:Cardiemems的早晨肺动脉压力测量最稳定,推荐用于?压力趋势监测

获取原文
       

摘要

Aims The CardioMEMS HF system is used to measure pulmonary artery (PA) pressures of patients with heart failure (HF). The goal of this study was to determine the impact of time in the daily PA pressure measurements, considering variance and influence of circadian rhythms on cardiovascular pathophysiology. Methods and results The study included 10 patients with HF with reduced ejection fraction (LVEF ?40%; New York Heart Association class III). Individual daily PA pressures were obtained by CardioMEMS sensors, per protocol, measured up to six times throughout the day, for a period of 5 days. Differences between variation of morning versus evening PA pressures were compared with Wilcoxon signed-rank test. Mean PA pressures (mPAP) increased from a morning value of 19.1?±?2?mm?Hg (8 am; mean?±?standard error of the mean [SEM]) to 21.3?±?2?mm?Hg late in the evening (11?pm; mean?±?SEM). Over the course of 5 days, evening mPAP exhibited a significantly higher median coefficient of variation than morning mPAP (14.9 (interquartile range [IQR] 7.6–21.0) and 7.0 (IQR 5.0–12.8) respectively; p?=?0.01). The same daily pattern of pressure variability was observed in diastolic (p?=?0.01) and systolic (p?=?0.04) pressures, with diastolic pressures being more variable than systolic at all time points. Conclusions Morning PA pressure measurements yield more stable values for observing PA trends. Patients should thus be advised to consistently perform their daily PA pressure measurements early in the morning. This will improve reliability and interpretation of the CardioMEMS management, indicating true alterations in the patient’s health status, rather than time-of-day-dependent variations.
机译:AIMS用于测量心力衰竭(HF)患者的肺动脉(PA)压力来测量肺动脉(PA)压力。本研究的目标是确定时间在每日PA压力测量中的影响,考虑到昼夜节律对心血管病理生理学的差异和影响。方法和结果该研究包括10例HF,其中射血分数降低(LVEF& 40%;纽约心结社III类)。每项协议的心肌传感器获得单独的每日PA压力,全天可达六次,为期5天。将早晨与晚间PA压力变化的差异与Wilcoxon签名秩检验进行比较。平均PA压力(MPAP)从19.1°的早晨值增加,±2?2?mm?hg(8m;平均值α±±±标准误差为21.3?±2?mm?hg晚上晚上(11?PM;平均值?±α滤光)。在5天的过程中,晚上MPAP分别表现出比早晨的MPAP(14.9(IQR] 7.6-21.0)和7.0(IQR 5.0-12.8)分别的明显更高的变异系数; P?= 0.01)。在舒张压(P?= 0.01)和收缩压(P?= 0.04)压力下观察到相同的压力变异图案,舒张压比所有时间点的收缩压更可变。结论早晨PA压力测量结果为观察PA趋势产生更稳定的值。因此,应建议患者始终清晨始终每天进行每日PA压力测量。这将提高心肌管理的可靠性和解释,表明患者健康状况的真正改变,而不是一时间依赖的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号