Adjusting external calibration intervals for auxiliary devices in testing laboratories according to intermediate checks results

机译:根据中间检查结果调整测试实验室中的辅助设备的外部校准间隔

获取原文
           

摘要

An unfortunate thing which is not desirable to happen in testing laboratories is that after working for weeks on tests, it is discovered that one of the used equipment is not accurate. Therefore, intermediate checking these devices from time to time is a reasonable way to reduce the risk of errors in laboratory work. The Central Laboratory for Environmental Quality Monitoring (CLEQM) performs intermediate check for testing equipment, besides the annual calibration done by an accredited body, to maintain confidence in its accuracy and precision between calibration intervals. This paper discusses the laboratory procedures to perform intermediate checks for electronic balances and micropipettes to comply with ISO/IEC:17025 requirements in the field of water quality testing. Intermediate check for balances was carried out according to OIML R111-2004 (E), while intermediate check for micropipettes was performed according to ISO 8655-6-2002. The results of intermediate checks and control charts were used to monitor equipment needs for recalibration. The results indicated that calibration of laboratory equipment is not necessarily neither to be annually nor at fixed time intervals. However, it should be based on the drift in measurements, which can be detected through regular intermediate check. Where results are tracked through control charts and in case that, it is drifted out of control limits, a nonconformance is reported, and appropriate corrective actions are taken including calibration if required. It is concluded that calibration of supporting equipment according to their intermediate check is more favorable than fixed time intervals due to the high degree of reliability of the generated test results in addition to reduced financial costs, which is expected to be 25% to 40% lower than the cost of calibration with fixed intervals.
机译:一个不幸的事情是在测试实验室中发生的是,在测试后工作几周后,发现其中一个用过的设备不准确。因此,从时刻中间检查这些设备是降低实验室工作中误差风险的合理方法。除了认证机构完成的年校准外,环境质量监测(CLEQM)的中央实验室还执行测试设备的中间检查,以维持校准间隔之间的准确性和精度的信心。本文讨论了对电子余额和微量化的中间检查进行实验室程序,以符合ISO / IEC:17025在水质测试领域的要求。根据 OIM1R111-2004(E)进行余量进行余量,同时根据

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号