首页> 外文期刊>Journal of orthopaedics and traumatology: official journal of the Italian Society of Orthopaedics and Traumatology >Intraoperative complications in total hip arthroplasty using a new cementless femoral implant (SP-CL)
【24h】

Intraoperative complications in total hip arthroplasty using a new cementless femoral implant (SP-CL)

机译:使用新的粘合性股骨植入物(SP-CL)的总髋关节形成术中的术中并发症

获取原文
           

摘要

Considering the excellent results already achieved in total hip arthroplasty (THA), new implants must be at least as safe as currently used implants and lead to longer survival. A new cementless femoral stem, SP-CL , has been introduced. The aim of this study is to evaluate intraoperative complications and assess the risk factors of THA with the SP-CL implant. All THA patients who were operated on using the SP-CL (LINK, Hamburg, Germany) implant between 2015 and 2018 were included in the analysis. Data were collected from medical records from national and hospital electronic databases. Radiological measurements were made from standard pre- and postoperative radiographs. A total of 222 THA were performed using the SP-CL implant. The average age of the patients was 56 years (14 77 years). There were 1 transient sciatic nerve injury, 1 acetabular fracture, and 11 (5.0%) intraoperative femoral fractures (IFF), of which 7 were treated with cerclage wire or titanium band during the operation while the other fractures were treated conservatively. None of the IFF patients were revised due to fracture during the follow-up period (one revision due to infection). The radiographic morphology of proximal femur was associated with increased risk of IFF (p = 0.02). The results of the current study demonstrate a 5% incidence of IFF when using the LINK SP-CL femoral stem in THA. The radiographic morphology of the proximal femur was an important predictor of IFF and should be assessed when using SP-CL . Level 4.
机译:考虑到总髋关节置换术(THA)已经实现的优异成果,新植入物必须至少与目前使用的植入物一样安全,并导致更长的存活率。已经引入了一种新的软泥股骨SP-CL。本研究的目的是评估术中并发症,并评估与SP-CL植入物的危险因素。在2015年和2018年间使用SP-CL(Link,Hamburg,Germany)植入的所有THA患者被包括在分析中。从国家和医院电子数据库的医疗记录收集数据。由标准和术后X线片进行放射学测量。使用SP-CL植入物进行总共222个Tha。患者的平均年龄为56岁(147七年)。有1个瞬态坐骨神经损伤,1个髋臼骨折和11个(5.0%)术中股骨骨折(IFF),其中在操作期间用CERCRACE线或钛带治疗,而其他骨折保守处理。由于随访期间的骨折(感染引起的一次修订),IFF患者因骨折而进行修订。近端股骨的射线照相形态与IFF的风险增加有关(P = 0.02)。目前研究的结果表明,当在THA中使用链路SP-CL股骨茎时,IFF的发生率为5%。近端股骨的射线照相形态是IFF的重要预测因子,应该在使用SP-CL时进行评估。等级4。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号