首页> 外文期刊>Journal of Korean medical science. >Comparison of Continuous ECG Monitoring by Wearable Patch Device and Conventional Telemonitoring Device
【24h】

Comparison of Continuous ECG Monitoring by Wearable Patch Device and Conventional Telemonitoring Device

机译:可穿戴贴片装置和传统遥测装置的连续ECG监控的比较

获取原文
           

摘要

BACKGROUND:Detection of arrhythmias is crucial for the treatment of cardiovascular diseases. However, conventional devices do not provide sufficient diagnostic accuracy while patients should suffer from bothersome diagnostic process. We sought to evaluate diagnostic capability and safety of the new adhesive electrocardiogram (ECG) monitoring device in patients who need ECG monitoring during admission.METHODS:We enrolled 10 patients who admitted to Seoul National University Bundang Hospital and required continuous ECG monitoring between October 31, 2019 and December 18, 2019. New adhesive ECG monitoring device and conventional ECG monitoring device were simultaneously applied to the patients and maintained for 48 hours. From each patient, 48 pairs of ECG signal were collected and analyzed by two cardiologists independently. Discrepancy of diagnosis and frequency of noise or signal loss were compared between the two devices.RESULTS:From analyzable ECG data, discrepancy of arrhythmia diagnosis was not observed between the two devices. Noise rate was higher in conventional ECG monitoring device (2.5% vs. 17.3%, P 0.001) and signal loss was not observed in new adhesive device while there was 9.4% of signal losses in conventional Holter recorder group. The new device was well-tolerated among 48 hours of monitoring period and no adverse event was observed.CONCLUSION:A newer adhesive ECG monitoring device demonstrated similar diagnostic accuracy compared to conventional ECG monitoring device.? 2020 The Korean Academy of Medical Sciences.
机译:背景:心律失常的检测对于治疗心血管疾病至关重要。然而,常规设备不提供足够的诊断准确性,而患者应患有麻痹性诊断过程。我们试图评估新的粘合心电图(ECG)监测装置的诊断能力和安全性,以期间需要ECG监测的患者。方法:我们注册了10月31日允许呼吁首尔国立大学的10名患者,并在10月31日之间进行持续的心电图监测, 2019年和2019年12月18日。新的粘合ECG监测装置和传统的ECG监测装置同时应用于患者并保持48小时。从每位患者,48对ECG信号被独立收集并由两个心脏病学家分析。在两个设备之间比较了诊断和噪声频率的差异或信号丢失。结果:来自可分析的ECG数据,两种设备之间未观察到心律失常诊断的差异。常规ECG监测装置中噪声率较高(2.5%对17.3%,P <0.001),在新的粘合装置中未观察到信号损耗,而传统的HOLTER记录仪组中的信号损耗占9.4%。新器件在监测期48小时内耐受良好,没有观察到不利事件。结论:与传统的ECG监测装置相比,更新的粘合剂ECG监测装置显示出类似的诊断准确性。? 2020韩国医学科学院。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号