Multiple Shots Averaging in Laser Flash Measurement

机译:激光闪光测量中的多次镜头平均

获取原文

摘要

The knowledge of the thermal properties of materials is a key resource for the correct design of industrial components. The laser flash method is a well-known procedure to determine the thermal diffusivity of a wide range of materials. However, in some cases there is the need of limiting the input power, or measuring materials with high thermal capacity, or investigating thick samples. Under these circumstances, the expected signal-to-noise ratio can reach extremely low values. To solve this issue, the existing laser flash apparatus has been improved, creating a new control and data acquisition system that is able to repeat multiple times the emission of the laser impulse and the measurement of the thermal response of the specimen. This gives the possibility of averaging several measurements, leading to a decrease of the noise. The experimental validation on different samples shows the feasibility of the proposed setup, that makes it possible to work with low power inputs.
机译:材料热特性的知识是工业部件正确设计的关键资源。激光闪光法是众所周知的方法,以确定各种材料的热扩散性。然而,在某些情况下,需要限制输入功率,或测量具有高热容量的材料,或者调查厚样品。在这种情况下,预期的信噪比可以达到极低的值。为了解决这个问题,已经提高了现有的激光闪光装置,创造了一种能够在多次发射激光脉冲和样本的热响应的测量来重复多次的新控制和数据采集系统。这给出了平均若干测量的可能性,从而降低了噪声。不同样本的实验验证显示了所提出的设置的可行性,使得可以使用低功耗输入。