首页> 外文期刊>JRSM Cardiovascular Disease >Accuracy of washout rate analysis for thallium-201 single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors: A phantom study
【24h】

Accuracy of washout rate analysis for thallium-201 single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors: A phantom study

机译:使用镉锌碲化镉探测器的铊-20单光子发射铊-2单光子发射灌注成像的洗涤率分析的准确性:Phantom研究

获取原文
       

摘要

Objective The aim of this study was to assess the accuracy of washout rate (WOR) analysis for thallium-201 chloride ( ~(201)Tl) single-photon emission computed tomography myocardial perfusion imaging data acquired using cadmium zinc telluride detectors and a myocardial phantom. Methods A myocardial phantom was injected with 10.5 MBq ~(201)Tl, and 10-min acquisitions were performed at 0, 24, 46, and 62?h to accommodate natural radioactive decay over time. Global myocardial WOR (global-WOR) and regional WOR (regional-WOR, left anterior descending artery [LAD], right coronary artery [RCA], and left circumflex artery [LCX]) were analyzed between 0 and 24?h (infarction model), 0 and 46?h (ischemia model), and 0 and 62?h (normal model), respectively. We compared the calculated radioactive decay-rate as a reference standard and phantom imaging WOR (phantom-WOR). Results Decay-rate versus phantom-WOR were 20.4% vs. 20.8% (global-WOR), 21.3% (LAD), 21.2% (RCA), and 19.7% (LCX) for the infarction model; 35.4% vs. 35.6% (global-WOR), 35.5% (LAD), 36.2% (RCA), and 35.2% (LCX) for the ischemia model; and 44.5% vs. 45.1% (global-WOR), 45.4% (LAD), 44.7% (RCA), and 43.5% (LCX) for the normal model. Conclusion WOR analysis for ~(201)Tl single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors is a reliable analysis method.
机译:目的本研究的目的是评估使用镉锌碲化镉探测器和心肌幻像获取的铊-20氯化铊 - 201氯化物(〜(201)T1)单光子发射计算断层摄影灌注成像数据的精度(〜(201)T1)的精度。方法使用10.5MBq〜(201)T1注射心肌体模,在0,24,46和62℃下进行10分钟,以适应天然放射性衰变随时间。在0到24℃(梗死模型),0和46?H(缺血模型)和0和62?H(正常模型)。我们将计算出的放射性衰减率与参考标准和幻影成像进行比较Wor(Phantom-Wor)。结果衰减率与幻影 - Wor为20.4%(全球 - 工作),21.3%(LAD),21.3%(RCA)和19.7%(LCX)的梗塞模型; 35.4%与35.6%(全球 - 工作),35.5%(LAD),36.2%(RCA)和35.2%(LCX)为缺血模型; 44.5%与45.1%(全球 - 工作),45.4%(LAD),44.7%(RCA)和43.5%(LCX)用于正常模型。结论〜(201)TL单光子发射的WOR分析计算断层摄影心肌灌注成像使用镉锌碲化镉探测器是一种可靠的分析方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号