Objectives This study aimed to investigate the reliability of 3‐dimensional (3D) ultrasound in screening for developmental dysplasia of the hip (DDH) by comparing the results with those of 2‐dimensional (2D) ultrasound. Methods One hundred five infants who were younger than 6 months were enrolled in this study. All of the infants underwent 2D and 3D ultrasound scanning for DDH by novices and experts, and the images were graded by a lead expert. The scanning time and image grades were analyzed by Studentt tests (P ?
机译:本研究的目标旨在通过将结果与二维(2D)超声的结果进行比较来研究三维(3D)超声筛查筛查发育不良(DDH)的可靠性。方法在本研究中注册了一百五个年龄较小的婴儿。通过新手和专家对DDH进行2D和3D超声扫描的所有婴儿,并且图像被铅专家评分。通过学生 T测试分析扫描时间和图像等级( p?<α.05)。通过腹部相关系数(ICC)评估新手和专家之间α角度测量的一致性。结果105名婴儿包括34名男孩和71名女孩。在2D扫描中,专家达成了关于正确诊断的同意,而在新手组中,41名婴儿误导了。在任一组中没有3D扫描的误诊。在新手组中,平均图像等级±SD为4.2?±1.3(2D超声)和8.1?±0.7(3D超声; P?<β.05)。在专家组中,平均图像等级为7.4?±1.0(2D超声)和8.2?±1.0(3D超声; p?<β.05)。 3D超声的等级中的群体之间没有统计学显着的差异( p = .83)。 3D超声波的扫描时间短于两个组中的2D超声(p≤1.05)中的2D超声。在新手组中,2D和3D超声结果之间的α角的ICC为0.34,并且在专家组中,它为0.92。当比较组和3D超声导致组的2D和3D超声导致时,ICC分别为0.35和0.84。结论三维超声需要较少的时间,并且表现出比2D超声更大的帧间间可靠性用于检测DDH。