首页> 外文期刊>Trauma Case Reports >Potentially serious adverse effects from application of a circumferential compression device for pelvic fracture: A report of three cases
【24h】

Potentially serious adverse effects from application of a circumferential compression device for pelvic fracture: A report of three cases

机译:潜在的严重不良反应,从盆腔骨折的圆周压缩装置的应用:三种情况的报告

获取原文
       

摘要

Pelvic circumferential compression devices (PCCDs) have gained wide acceptance in the management of patients with pelvic fracture. These devices are considered safe due to their noninvasive nature and significant hazards associated with the use of PCCDs have not been reported previously. However, we present herein the cases of three patients who received PCCD application and eventually developed major complications presumably caused by PCCDs. As a result, one patient developed surgical site infection following internal fixation and required several debridements. Another patient ended up with a walking disability. The remaining patient eventually died from exsanguination following application of the PCCD. Clinicians should be aware of the potential for deleterious effects, including bladder rupture, muscle necrosis, and vessel injuries. In particular, application for acetabular fractures and prolonged application of PCCDs should be avoided.
机译:盆腔圆周压缩装置(PCCD)在盆腔骨折患者的管理方面取得了广泛的认可。由于它们之前,这些装置被认为是安全的,并且之前尚未报道与使用PCCD相关的显着危险。然而,我们在此呈现3例接受PCCD申请的患者,最终产生了由PCCD引起的主要并发症。因此,在内部固定后,一名患者发育了手术部位感染,需要几个清血统。另一位患者以行走的残疾最终结束。在施用PCCD后,剩余的患者最终从放血中死亡。临床医生应该了解有害影响的可能性,包括膀胱破裂,肌肉坏死和血管受伤。特别是,应避免患有髋臼骨折和延长施用PCCD的应用。