FT-IR Spectroscopy for the Detection of Diethylene Glycol (DEG) Contaminant in Glycerin-Based Pharmaceutical Products and Food Supplements

机译:FT-IR光谱检测甘油基药品和食品补充剂中的二乙二醇(DEG)污染物

获取原文
获取外文期刊封面目录资料

摘要

Identification and determination of diethylene glycol (DEG) in glycerin-based products was successfully achieved using FT-IR spectroscopy. Studied samples included 0.5% to 20% by mass DEGspiked into cough syrup, two paracetamol syrup formulations, and two food supplements. The characteristic DEGwavenumbers at 881?cm -1 and 1083?cm -1 were used for its quantitative determination in the studied samples. A very good accuracy in determining the DEG fraction was achieved with a mean error% of ±2.02% to ±7.69% upon using the corrected absorbance at 881?cm -1 . The corrected absorbance at 1083?cm -1 band was used in the case of paracetamol formulations and resulted in a mean error% ranging from ±2.50% to ±10.28%. The values of limit of detection of the current method ranged from 0.051% to 0.068% DEG for all studied samples.
机译:使用FT-IR光谱成功地实现了甘油基产物中二乙二醇(DEG)的鉴定和测定。研究样品包括0.5%至20%受到质量的咳嗽糖浆,两种扑热息痛糖浆制剂和两种食品补充剂。在881〜Cm -1和1083Ω·cm -1处的特征Degwavenumber用于研究所研究的样品中的定量测定。在使用881Ωcm-1的校正吸光度时,在均衡误差±2.02%至±7.69%的平均误差%达到非常好的精度。在扑热酰胺制剂的情况下使用1083Ω·cm -1带的校正吸光度,并导致平均误差%范围为±2.50%至±10.28%。目前方法的检测限值范围为所有研究的样品的0.051%至0.068%。