首页> 外文期刊>Clinical ophthalmology >Comparison of Biometry Measurements Using Standard Partial Coherence Interferometry versus New Scheimpflug Tomography with Integrated Axial Length Capability
【24h】

Comparison of Biometry Measurements Using Standard Partial Coherence Interferometry versus New Scheimpflug Tomography with Integrated Axial Length Capability

机译:使用标准部分连贯干涉测量法与集成轴向长度能力的新型剖面性断层扫描的生物谱测量比较

获取原文
           

摘要

Purpose: To compare biometry measurements obtained by a partial interferometer biometer (IOLMaster 500) to the new Scheimpflug tomography with an integrated axial length biometer module (Pentacam AXL). Patients and Methods: Cataract patients who underwent biometric measurements with the IOL Master 500 and the Pentacam AXL from July to November 2017 were enrolled in this study. Comparisons were performed for axial length (AL), keratometry (K), and anterior chamber depth (ACD). The Pearson correlation coefficient and the 95% limits of agreement (LoA) were calculated. Paired Student’s t -tests and Bland-Altman plots were used to assess the differences between devices. Results: One hundred and sixty-six eyes of 92 patients were analyzed. There were no statistically significant differences in AL (p=0.558) or flat K (p=0.196) values between the IOL Master 500 and Pentacam AXL measurements. Statistically significant differences were found between the two devices with respect to steep K, ACD, and mean K measurements (p 0.001). Conclusion: Both devices provided similar measurements of AL and flat K, though there were statistically significant differences in ACD, steep K, and mean K measurements.
机译:目的:将部分干涉仪生物仪(IOLMASTER 500)与集成轴向长度的生物计模块(PENTACAM AXL)进行新的Scheimpful断层扫描进行比较使用部分干涉仪生物仪(Iolmaster 500)。患者和方法:从2017年7月到2017年7月到2017年7月到2017年7月到2017年11月的生物识别测量的白内障患者。对轴向长度(Al),腔腔谱(K)和前房深度(ACD)进行比较。计算Pearson相关系数和协议(LOA)的95%限制。配对的学生的T -Tests和Bland-Altman情节用于评估设备之间的差异。结果:分析了92例患者的一百六十六次眼睛。在IOL主500和Pentacam AXL测量之间,Al(p = 0.558)或扁平K(p = 0.196)值没有统计学上显着的差异。在两个装置相对于陡峭K,ACD和平均值测量之间发现了统计学显着的差异(P <0.001)。结论:两种装置提供了Al和扁平K的类似测量,尽管ACD,陡峭K和平均值测量存在统计学上存在统计学意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号