Ex vivo biomechanical comparison of 2.7?mm string-of-pearl plate versus screw/wire/Polymethylmethacrylate composite fixation and 2.7?mm veterinary acetabular plate for repair of simulated canine acetabular fractures

机译:珠子串柱的2.7毫米/聚甲基丙烯酸甲酯复合固定和2.7毫米兽药骨架,用于修复模拟犬髋臼骨折

获取原文

摘要

Acetabular fractures comprise 12–30% of canine pelvic fractures and require accurate anatomic reduction and rigid stability to ensure proper healing and minimize future osteoarthritis. Many techniques have been used to repair these fractures, with common techniques including veterinary acetabular plates or use of screw/wire/polymethylmethacrylate constructs. String-of-Pearl? plating systems have also been used clinically but there is a lack of research supporting their use for these fractures. The purpose of this study was to compare fracture reduction accuracy, biomechanical characteristics, and mode of failure between String-of-Pearls?, veterinary acetabular plates, screw/wire/polymethylmethacrylate constructs in a simulated, ex-vivo acetabular fracture model. We hypothesized that the String-of-Pearls? constructs would have equivalent or greater mechanical properties and reduction compared to the other constructs. The mean craniocaudal acetabular diameter before fixation (mean 25.2?mm; range 20?mm – 30.1?mm) was not significantly different from after fixation (mean 23.9?mm; range 20?mm – 28.3?mm) for any fixation method. Comparison of reduction scores between groups revealed no significant differences. No significant differences were noted for cyclical displacement or stiffness. There was significant difference with superior failure load of String-of-Pearls? compared to screw/wire/polymethylmethacrylate in the 75th percentile of animal weight (P?=?0.0021), and superior failure load of String-of-Pearls? compared to veterinary acetabular plates in the 50th (P?=?0.0232) and 75th percentiles (P?=?0.0058). Stiffness of the String-of-Pearls? construct was significantly greater than the veterinary acetabular plate construct (P?=?0.0417). For ultimate load, String-of-Pearls? constructs were significantly greater than screw/wire/polymethylmethacrylate (P?=?0.0331) and veterinary acetabular plates (P?=?0.0218). Although the ease of application for the String-of-Pearls? implant was subjectively better than other implants, no significant differences were found in fracture reduction scores. The String-of-Pearls? constructs were stiffer than veterinary acetabular plates and exhibited greater failure and ultimate loads compared to veterinary acetabular plates and screw/wire/polymethylmethacrylate fixations. The String-of-Pearls? implant appears to be a suitable fixation choice for simple canine acetabular fractures.
机译:髋臼骨折包括12-30%的犬骨盆骨折,需要准确的解剖结构和刚性稳定性,以确保适当的愈合和最小化未来的骨关节炎。许多技术已被用于修复这些裂缝,具有普通技术,包括兽医髋臼板或使用螺丝/线/聚甲基丙烯酸甲酯构建体。珍珠串?电镀系统也在临床上使用,但缺乏支持它们对这些骨折的使用。本研究的目的是比较骨折的骨折精度,生物力学特性和失效模式,兽医髋臼板,螺旋/线/聚甲基丙烯酸甲酯构建体在模拟的前体内髋臼骨折模型中。我们假设珍珠串?与其他构建体相比,构造具有等同的或更大的机械性能和还原。固定前(平均25.2ΩΩmm)的平均颅骨髋臼直径与固定后(平均23.9μm;范围为20×mm - 28.3Ω·mm)没有显着不同。组之间的减少评分的比较显示没有显着差异。对于循环位移或刚度没有显着差异。珍珠串的卓越故障负荷有显着差异吗?与75百分位的动物重量(p?= 0.0021)中的螺丝/丝/聚甲基丙烯酸甲酯相比,以及珠粒串的卓越故障负荷?与50th(p?= 0.0232)和第75百分位数(p?= 0.0058)相比兽医髋臼板。珍珠柱的刚度?构建体显着大于兽医髋臼板构建体(P?= 0.0417)。对于终极载荷,珍珠串?构建体显着大于螺丝/线/聚甲基丙烯酸甲酯(P?= 0.0331)和兽医髋臼板(P?= 0.0218)。虽然易于珍珠的申请?植入物的主观优于其他植入物,在断裂分数中没有发现显着差异。珍珠串?与兽医髋臼板和螺丝/线/聚甲基丙烯酸甲酯固定相比,构造比兽医髋臼板更硬,并且与兽医髋臼板和螺丝/线/聚甲基丙烯酸甲酯固定表现出更大的失效和最终载荷。珍珠串?植入物似乎是简单的犬髋臼骨折的合适固定选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号