Retinal-image quality and contrast sensitivity function in eyes with epiretinal membrane: a cross-sectional observational clinical study

机译:表位膜的眼睛视网膜图像质量和对比度敏感功能:横截面观察临床研究

获取原文
           

摘要

To investigate the effect of idiopathic epiretinal membrane (ERM) on the retinal-image quality and psychophysical contrast sensitivity function (CSF). Forty-four subjects with diagnosis of idiopathic unilateral ERM were enrolled in this cross-sectional observational clinical study. The fellow unaffected eyes were set as the control group. For retinal-image quality assessment, an Optical Quality Analysis System (OQAS) based on double-pass technique was used to evaluate objective scatter index (OSI) and Strehl ratio. For visual performance, the CSF under photopic condition was measured. For retinal-image quality, the result of double-pass device revealed a significant lower Strehl ratio and larger OSI in the ERM eyes compared to the fellow eyes (all P??0.05). For visual performance, the CSF at all spatial frequencies under photopic condition were also significantly degraded in the ERM eyes compared to the fellow eyes (all P??0.05). For the ERM eyes, the reduction of Strehl ratio and CSF was 29.41 and 54.39%, respectively, and the increase of OSI was 164.10% compared to the fellow eyes. Besides, BCVA significantly correlated to the total CSF (ERM eyes, r?=???0.53, P??0.001; the fellow eyes, r?=???0.467, P?=?0.002) and Strehl ratio (ERM eyes, r?=???0.485, P?=?0.001; the fellow eyes, r?=???0.311, P?=?0.043) in both of the ERM and the fellow eyes. Eyes affected with ERM showed poorer retinal-image quality and visual performance than the normal eyes. Retinal-image quality measured by OQAS based on double-pass technique could be useful for assessing the retinal-image quality for ERM-affected eyes, in which retinal scattering was significantly increased.
机译:探讨特发性脑膜膜膜(ERM)对视网膜图像质量和心理物理对比敏感功能(CSF)的影响。在这种横断面观察临床研究中注册了诊断特发性单侧ERM的四十四个受试者。同胞不受影响的眼睛被设定为对照组。对于视网膜图像质量评估,使用基于双通技术的光学质量分析系统(OQAS)来评估客观散射指数(OSI)和STHLL比率。为了进行视觉性能,测量了光学条件下的CSF。对于视网膜图像质量,双通装置的结果显示出与同伴眼睛(所有P?<0.05)相比,ERM眼中的显着低的斯特勒比和较大的OSI。对于视觉性能,与同伴(所有P?<0.05)相比,光学频率下的所有空间频率下的CSF也显着降低。对于ERM的眼睛,STREHL比和CSF的降低分别为29.41和54.39%,与同伴的眼睛相比,OSI的增加为164.10%。此外,BCVA与总CSF(ERM Eyes)显着相关(ERM Eyes,R ??? 0.53,P?<0.001;同伴的眼睛,R?= ??? 0.467,P?= 0.002)和STREHL比率(ERM眼睛,r ??? 0.485,p?= 0.001;同胞,r?= ??? 0.311,p?= 0.043),两者和同胞的眼睛。对ERM影响的眼睛显示出较差的视网膜图像质量和视觉性能而不是正常的眼睛。基于双通技术的OQAS测量的视网膜图像质量可用于评估ERM受影响的视网膜图像质量,其中视网膜散射显着增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号