首页> 外文期刊>The Journal of Engineering >Study on the performance of X-ray machines used in K-fluorescent devices
【24h】

Study on the performance of X-ray machines used in K-fluorescent devices

机译:K-荧光装置中使用的X射线机的性能研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Since Roentgen discovered X-rays, X-rays have been widely used in medical diagnosis and industrial non-destructive testing. With the improvement of the performance of aviation technology and instrumentation, higher requirements have been set for the calibration of detectors. X-ray machine performance is very important for calibration. Tube voltage, tube current, and inherent filtration are important parameters of the X-ray machine and directly affect the image imaging quality and radiation dose. Here, the voltage performance of the X-ray machine was measured using HPGe detector. Measure the tube voltage of the X-ray machine using the end of the spectrum method, it is shown that the performance of the X-ray machine shows the tube voltage within +/- 2%. Measurement of the linearity of the X-ray tube current and current stability using the PTW30013 ionisation chamber. The relative deviation of the actual measured tube current is 0.08%. The inherent filtration of the X-ray machine was measured by the half-value layer method, and the inherent filtration was 0.058mm Al. The inherent filtration of the K-fluorescent device was measured to be 1.487mm Al.
机译:由于Roentgen发现X射线,因此X射线已广泛用于医学诊断和工业无损检测。随着航空技术和仪器性能的提高,已经为探测器的校准设定了更高的要求。 X射线机性能对于校准非常重要。管电压,管电流和固有过滤是X射线机的重要参数,直接影响图像成像质量和辐射剂量。这里,使用HPGE检测器测量X射线机的电压性能。使用光谱法的结束测量X射线机的管电压,表明X射线机的性能显示在+/- 2%内的管电压。使用PTW30013电离室测量X射线管电流的线性和电流稳定性。实际测量管电流的相对偏差为0.08%。通过半值层法测量X射线机的固有过滤,固有过滤为0.058mm。测量K-荧光装置的固有过滤为1.487mm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号