Verification of Radiation Fluence using Stack Film in HDR Brachytherapy with Heterogeneity Algorithm

机译:HDR近距离放射治疗中使用堆叠膜的异质性算法验证辐射通量

获取原文
获取外文期刊封面目录资料

摘要

It is essential that Quality Assurance (QA) need to be in line with progresses in HDR brachytherapy treatment planning and delivery to ensure an appropriate level of dosimetric accuracy and quality.Aim: To verify radiation fluence using stack films in HDR brachytherapy with heterogeneity algorithm.Materials and Methods: Gafchromic EBT2 films were stacked in different experimental setup and each film acted as dosimeter to validate AcurosTMBV algorithm in HDR brachytherapy. These films were read using Epson expression 10000 XL flatbed scanner, calibrated against the standard value and analysed with Film QA Pro 2015 software. Different criterion were set for gamma analysis which include Dose Difference (DD)/Distance-To-Agreement (DTA); 2%/1 mm, 5%/1 mm, 10%/1 mm.Results: In order to validate the point doses calculated using model based dose calculation algorithm (AcurosTM BV), a virtual phantom created in the TPS, which agreed with MC based calculation mostly within 2%. The results of each plane were analysed in all the experimental setup. In the gamma pass criteria of 2% and 1 mm, 92% passed, in 5% and 1 mm criteria 96% passed and with 10% and 1 mm criteria, almost 100% passing rate was achieved.Conclusion: Methods provide a comprehensive verification for commissioning the MBDCA-AcurosTM BV and are recommended to use stack film for QA program in HDR brachytherapy.
机译:质量保证(QA)必须与HDR近距离放射治疗计划和交付的进展保持一致,以确保适当水平的剂量学准确性和质量。目的:通过在HDR近距离放射治疗中使用堆叠膜来验证辐射通量材料和方法:将全色EBT2薄膜堆叠在不同的实验装置中,每片薄膜均充当剂量计,以验证AcurosTMBV算法在HDR近距离放射治疗中的有效性。使用Epson expression 10000 XL平板扫描仪读取这些胶片,针对标准值进行校准,并使用Film QA Pro 2015软件进行分析。为伽玛分析设置了不同的标准,包括剂量差异(DD)/协议一致距离(DTA);结果:为了验证使用基于模型的剂量计算算法(AcurosTM BV)(在TPS中创建的虚拟体模)计算的点剂量,请验证2%/ 1 mm,5%/ 1 mm,10%/ 1 mm。这与基于MC的计算结果相吻合,大多在2%以内。在所有实验设置中分析了每个平面的结果。在2%和1 mm的伽玛通过标准中,通过率为92%,在5%和1 mm的准则中为96%通过,而在10%和1 mm的准则中,则达到了近100%的通过率。结论:方法提供对调试MBDCA-AcurosTM BV进行了全面的验证,建议在HDR近距离放射治疗中将堆叠膜用于QA程序。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号