首页> 外文期刊>Journal of Clinical Microbiology >Comparison of Automated Antimicrobial Susceptibility Testing Systems To Detect mecC-Positive Methicillin-Resistant Staphylococcus aureus
【24h】

Comparison of Automated Antimicrobial Susceptibility Testing Systems To Detect mecC-Positive Methicillin-Resistant Staphylococcus aureus

机译:自动化的药敏试验系统检测耐mecC的耐甲氧西林金黄色葡萄球菌的比较

获取原文
           

摘要

A total of 111 MRSA isolates of human and animal origin collected in different European countries (England, Scotland, Denmark, and France) confirmed as mecC positive by PCR (2) were included in this study. We compared the AST results obtained for oxacillin and cefoxitin with (i) the BD Phoenix PMIC/ID-60 panel (BD Diagnostic Systems, Franklin Lakes, NJ), (ii) the Vitek 2 AST-P581 card (bioMérieux, Marcy l'Etoile, France), and (iii) the MicroScan Pos MIC type 31 panel (Beckman Coulter, Brea, CA). In accordance with CA-SFM/EUCAST and CLSI guidelines, the following breakpoints were used to assign methicillin resistance: 2 mg/liter for oxacillin and 4 mg/liter for cefoxitin.
机译:本研究包括在不同欧洲国家(英格兰,苏格兰,丹麦和法国)收集的111种人和动物MRSA分离株,经PCR证实为mecC阳性(2)。我们将奥沙西林和头孢西丁的AST结果与(i)BD Phoenix PMIC / ID-60面板(BD Diagnostic Systems,Franklin Lakes,NJ),(ii)Vitek 2 AST-P581卡(bioMérieux,Marcy l'法国Etoile)和(iii)MicroScan Pos MIC类型31面板(加利福尼亚州布雷亚的贝克曼库尔特)。根据CA-SFM / EUCAST和CLSI指南,使用以下断点确定甲氧西林的耐药性:奥沙西林> 2 mg / L,头孢西丁> 4 mg / L。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号