首页> 外文期刊>Journal of Clinical Microbiology >Comparison of Specimen Processing and Nucleic Acid Extraction by the Swab Extraction Tube System versus the MagNA Pure LC System for Laboratory Diagnosis of Herpes Simplex Virus Infections by LightCycler PCR
【24h】

Comparison of Specimen Processing and Nucleic Acid Extraction by the Swab Extraction Tube System versus the MagNA Pure LC System for Laboratory Diagnosis of Herpes Simplex Virus Infections by LightCycler PCR

机译:用拭子提取管系统和MagNA纯LC系统对标本处理和核酸提取的比较,用于通过LightCycler PCR实验室诊断单纯疱疹病毒感染

获取原文
       

摘要

A total of 563 specimens (234 dermal and 329 genital swabs) from patients suspected of having herpes simplex virus (HSV) infections were processed using two different extraction methods (the MagNA Pure LC system and the swab extraction tube system [SETS]); HSV DNA was amplified by LightCycler PCR. HSV DNA was detected in 157 of 563 specimens (27.9%) processed by the MagNA Pure LC system and in 179 of 563 specimens (31.8%) processed by SETS (P < 0.0001). There was no specimen processed by the MagNA Pure LC extraction method that was positive only for HSV DNA. Of 157 specimens positive by both methods, HSV DNA copy levels were higher (using cycle crossover points [cycle threshold {CT}]) with SETS (mean CT, 25.9 cycles) than with the MagNA Pure LC system (mean CT, 32.0 cycles) (P < 0.0001). The time to process 32 samples was longer with the MagNA Pure LC extraction system (90 min) than with SETS (35 min). HSV DNA extraction using SETS is faster, less expensive, and more sensitive than the MagNA Pure LC system and could replace the latter for the laboratory diagnosis of HSV infections using LightCycler PCR.
机译:使用两种不同的提取方法(MagNA Pure LC系统和拭子提取管系统[SETS])对来自怀疑患有单纯疱疹病毒(HSV)感染的患者的563个标本(234个皮肤拭子和329个生殖器拭子)进行了处理; HSV DNA通过LightCycler PCR扩增。通过MagNA Pure LC系统处理的563个样本中的157个(27.9%)和通过SETS处理的563个样本(31.8%)中的179个中检测到HSV DNA( P <0.0001)。没有通过MagNA Pure LC提取方法处理过的标本仅对HSV DNA呈阳性的标本。在这两种方法均为阳性的157个标本中,HSV DNA复制水平较高(使用循环交叉点[循环阈值{ C T }]),且SETS(均值 C T ,25.9个周期)比使用MagNA Pure LC系统(平均 C T ,32.0个周期)( P <0.0001)。 MagNA Pure LC萃取系统处理样品的时间(90分钟)比SETS处理样品的时间(35分钟)更长。与MagNA Pure LC系统相比,使用SETS提取HSV DNA更快,更便宜且更灵敏,并且可以代替后者用于通过LightCycler PCR在实验室诊断HSV感染。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号