首页> 外文期刊>The Journal of Bone and Joint Surgery. British VolumecBritish Orthopaedic Association , Australian Orthopaedic Association , Canadian Orthopaedic Association . . . [et al] >Validation of primary metal-on-metal hip arthroplasties on the National Joint Registry for England, Wales and Northern Ireland using data from the London Implant Retrieval Centre
【24h】

Validation of primary metal-on-metal hip arthroplasties on the National Joint Registry for England, Wales and Northern Ireland using data from the London Implant Retrieval Centre

机译:使用伦敦植入物检索中心的数据,在英格兰,威尔士和北爱尔兰的国家联合登记处对主要的金属对金属髋关节置换术进行验证

获取原文
           

摘要

Arthroplasty registries are important for the surveillance of joint replacements and the evaluation of outcome. Independent validation of registry data ensures high quality. The ability for orthopaedic implant retrieval centres to validate registry data is not known. We analysed data from the National Joint Registry for England, Wales and Northern Ireland (NJR) for primary metal-on-metal hip arthroplasties performed between 2003 and 2013. Records were linked to the London Implant Retrieval Centre (RC) for validation. A total of 67?045 procedures on the NJR and 782 revised pairs of components from the RC were included. We were able to link 476 procedures (60.9%) recorded with the RC to the NJR successfully. However, 306 procedures (39.1%) could not be linked. The outcome recorded by the NJR (as either revised, unrevised or death) for a primary procedure was incorrect in 79 linked cases (16.6%). The rate of registry-retrieval linkage and correct assignment of outcome code improved over time. The rates of error for component reference numbers on the NJR were as follows: femoral head category number 14/229 (5.0%); femoral head batch number 13/232 (5.3%); acetabular component category number 2/293 (0.7%) and acetabular component batch number 24/347 (6.5%). Registry-retrieval linkage provided a novel means for the validation of data, particularly for component fields. This study suggests that NJR reports may underestimate rates of revision for many types of metal-on-metal hip replacement. This is topical given the increasing scope for NJR data. We recommend a system for continuous independent evaluation of the quality and validity of NJR data.
机译:关节置换登记对于监视关节置换和评估结局非常重要。独立验证注册表数据可确保高质量。骨科植入物检索中心验证注册表数据的能力尚不清楚。我们分析了英格兰,威尔士和北爱尔兰国家联合注册局(NJR)的数据,该数据用于2003年至2013年之间进行的主要金属对金属髋关节置换术。记录与伦敦植入物检索中心(RC)关联以进行验证。总共对NJR进行了67-045个程序,并从RC修改了782对组件。我们能够成功地将RC记录的476个程序(占60.9%)链接到NJR。但是,无法链接306个过程(占39.1%)。 NJR记录的主要手术结局(修订,未修订或死亡)在79例相关案件中不正确(16.6%)。随着时间的流逝,注册表检索链接和正确分配结果代码的比率有所提高。 NJR上的组件参考编号的错误率如下:股骨头类别编号14/229(5.0%);股骨头批号13/232(5.3%);髋臼组件类别编号2/293(0.7%)和髋臼组件批次编号24/347(6.5%)。注册表检索链接提供了一种新颖的方法来验证数据,尤其是对于组件字段。这项研究表明,NJR报告可能会低估许多类型的金属对金属髋关节置换术的翻修率。鉴于NJR数据的范围越来越大,这是热门话题。我们建议使用一种连续不断地独立评估NJR数据的质量和有效性的系统。
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号