首页> 外文期刊>The Journal of Bone and Joint Surgery. British VolumecBritish Orthopaedic Association , Australian Orthopaedic Association , Canadian Orthopaedic Association . . . [et al] >Qualitative α-defensin test (Synovasure) for the diagnosis of periprosthetic infection in revision total joint arthroplasty
【24h】

Qualitative α-defensin test (Synovasure) for the diagnosis of periprosthetic infection in revision total joint arthroplasty

机译:定性α-防御素测试(Synovasure)在翻修全关节置换术中诊断假体周围感染

获取原文
获取外文期刊封面目录资料

摘要

AimsThe diagnosis of periprosthetic joint infection (PJI) remains demanding due to limitations of all the available diagnostic tests. The synovial fluid marker, α-defensin, is a promising adjunct for the assessment of potential PJI. The purpose of this study was to investigate the qualitative assessment of α-defensin, using Synovasure to detect or exclude periprosthetic infection in total joint arthroplasty.Patients and MethodsWe studied 50 patients (28 women, 22 men, mean age 65 years; 20 to 89) with a clinical indication for revision arthroplasty who met the inclusion criteria of this prospective diagnostic study. The presence of α-defensin was determined using the qualitative Synovasure test and compared with standard diagnostic methods for PJI. Based on modified Musculoskeletal Infection Society (MSIS) criteria, 13?cases were categorised as septic and 36 as aseptic revisions. One test was inconclusive.ResultsThe Synovasure test achieved a sensitivity of 69% and a specificity of 94%. The positive and negative likelihood ratios were 12.46 and 0.33, respectively. A good diagnostic accuracy for PJI, with an area under the curve of 0.82, was demonstrated. Adjusted p-values using the?method of Hochberg showed that Synovasure is as good at diagnosing PJI as histology (p = 0.0042) and bacteriology with one positive culture (p = 0.0327).ConclusionWith its ease of use and rapid results after approximately ten minutes, Synovasure may be a useful adjunct in the diagnosis of PJI.
机译:目的由于所有可用诊断测试的局限性,假体周围关节感染(PJI)的诊断仍然很困难。滑液标记物α-防御素是评估潜在PJI的有希望的辅助手段。这项研究的目的是研究使用Synovasure在全关节置换术中检测或排除假体周围感染的情况下对α-防御素的定性评估。患者和方法我们研究了50例患者(28名女性,22名男性,平均年龄65岁; 20至89岁) ),具有符合这项前瞻性诊断研究纳入标准的翻新置换术的临床指征。使用定性Synovasure测试确定α-防御素的存在,并将其与PJI的标准诊断方法进行比较。根据改良的肌肉骨骼感染学会(MSIS)标准,将13例病例归为败血症,将36例病例归为无菌菌。一项测试尚无定论。结果Synovasure测试的敏感性为69%,特异性为94%。正和负似然比分别为12.46和0.33。 PJI具有良好的诊断准确性,其曲线下面积为0.82。使用Hochberg方法对p值进行调整后,发现Synovasure与组织学(p = 0.0042)和细菌学与一种阳性培养物(p = 0.0327)一样,在诊断PJI方面同样出色。结论易于使用,约十分钟后即可快速获得结果。 ,Synovasure可能是诊断PJI的有用辅助手段。
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号