We report the effect of padding on the efficiency of the pneumatic tourniquet for the upper limb. Varying thicknesses of two commercially-available types of orthopaedic padding (Cellona and Velband) were applied to the arms of 20 volunteers, with three pressure transducers placed directly beneath the padding. A tourniquet was positioned over the padding and inflated to 220 mmHg.Significant reductions in the transmitted pressure were recorded from the transducers with both padding materials. With eight layers of padding, reductions in pressure of 13% (1% to 26%) and 18% (7% to 35%) were seen with Cellona and Velband, respectively. The reduction in pressure with Velband padding correlated with increasing arm circumference (Pearson’s correlation coefficient 0.711, p < 0.001).Studies to date have examined how arm circumference affects the required tourniquet inflation pressure. Our study is the first to investigate the effect of the padding and the findings suggest that using more than two layers results in a significant reduction in the transmitted pressure.
机译:我们报告了填充对上肢气动止血带效率的影响。将两种厚度不同的两种市售矫形垫(Cellona和Velband)应用于20名志愿者的手臂,并在垫的正下方放置三个压力传感器。将止血带置于衬垫上方并充气至220 mmHg。使用两种衬垫材料从换能器记录的传输压力显着降低。使用八层填充物,Cellona和Velband的压力分别降低了13%(1%至26%)和18%(7%至35%)。 Velband衬垫的压力降低与臂围增加有关(Pearson相关系数0.711,p <0.001)。迄今为止的研究已经研究了臂围如何影响止血带充气压力。我们的研究是第一个研究填充效果的研究,研究结果表明,使用两层以上的材料可以显着降低传输压力。