Precision of the measurements of periprosthetic bone mineral density in hips with a custom-made femoral stem

机译:使用定制的股骨柄测量髋关节假体周围骨矿物质密度的精度

获取原文

摘要

Our aim was to determine the precision of the measurements of bone mineral density (BMD) by dual-energy x-ray absorptiometry in the proximal femur before and after implantation of an uncemented implant, with particular regard to the significance of retro- and prospective studies.We examined 60 patients to determine the difference in preoperative BMD between osteoarthritic and healthy hips. The results showed a preoperative BMD of the affected hip which was lower by a mean of 4% and by a maximum of 9% compared with the opposite side. In addition, measurements were made in the operated hip before and at ten days after operation to determine the effect of the implantation of an uncemented custom-made femoral stem. The mean increase in the BMD was 8% and the maximum was 24%. Previous retrospective studies have reported a marked loss of BMD on the operated side.The precision of double measurements using a special foot jig showed a modified coefficient of variation of 0.6% for the non-operated side in 15 patients and of 0.6% for the operated femur in 20 patients.The effect of rotation on the precision of the measurements after implantation of an uncemented femoral stem was determined in ten explanted femora and for the operated side in ten patients at 10° rotation and in 20 patients at 30° rotation. Rotation within 30° influenced the precision in studies in vivo and in vitro by a mean of 3% and in single cases in up to 60%.Precise prediction of the degree of loss of BMD is thus only possible in prospective cross-sectional measurements, since the effect of the difference in preoperative BMD, as well as the apparent increase in BMD after implantation of an uncemented stem, is not known from retrospective studies. The DEXA method is a reliable procedure for determining periprosthetic BMD when positioning and rotation are strictly controlled.
机译:我们的目的是确定在未骨水泥植入物植入前后,通过双能x线骨密度仪对股骨近端进行骨矿物质密度(BMD)测量的准确性,特别是回顾性研究和前瞻性研究的重要性我们检查了60名患者,以确定骨关节炎和健康髋部术前BMD的差异。结果显示,患侧髋关节的术前BMD与对侧相比平均降低了4%,最大降低了9%。另外,在手术前和手术后十天在手术的髋部进行测量,以确定未骨水泥定制的股骨柄植入的效果。 BMD平均增加8%,最大增加24%。以前的回顾性研究报告了手术侧BMD的明显损失。使用特殊的脚夹具进行两次测量的精度显示,在15例患者中,非手术侧的修正变异系数为0.6%,非手术侧的修正变异系数为0.6% 20例患者中的股骨。在10例植入的股骨中以及10例10°旋转的患者和20例30°旋转的患者中,测定了旋转对未固定股骨柄植入后测量精度的影响。 30°范围内的旋转会平均影响体内和体外研究的精度3%,在单个情况下则高达60%。因此,只有在预期的横截面测量中才能准确预测BMD的丢失程度,由于回顾性研究尚不清楚术前BMD差异的影响以及植入非骨水泥茎后BMD明显增加的影响。当严格控制位置和旋转时,DEXA方法是确定假体周围骨密度的可靠方法。
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号