【24h】

Assessing the dog as a model for human total hip replacement

机译:将狗评估为人类全髋关节置换的模型

获取原文
   

获取外文期刊封面封底 >>

       

摘要

Post-mortem retrieval of canine, cemented femoral components was analysed to assess the performance of these implants in the dog as a model for human total hip replacement (THR). Mechanical testing and radiological analysis were performed to determine the stability of the implant and the quality of the cement. Thirty-eight implants from 29 dogs were retrieved after time intervals ranging from 0.67 to 11.67 years. The incidence of aseptic loosening was 63.2%, much higher than in human patients (6% in post-mortem studies). Failure of the femoral implants began with debonding at the cement-metal interface, similar to that in implants in man. The incidence of aseptic loosening was much lower in bilateral than in unilateral implants. Significant differences were observed for three different designs of implant. While the dog remains the animal model of choice for THR, results from this study provide insight into interspecies differences in the performance of implants. For example, the performance of THR in dogs should be compared with that in young rather than in elderly human patients.
机译:对尸体,骨水泥成分的尸体取回进行分析,以评估这些植入物在狗中作为人类全髋关节置换(THR)模型的性能。进行了机械测试和放射学分析,以确定植入物的稳定性和水泥的质量。在0.67至11.67年的时间间隔后,从29只狗中取出了38个植入物。无菌性松动的发生率为63.2%,远高于人类患者(验尸研究中为6%)。股骨植入物的破坏始于在水泥-金属界面处的脱粘,类似于人体内植入物。双侧无菌性松动的发生率远低于单侧植入物。观察到三种不同设计的植入物存在显着差异。尽管狗仍然是THR选择的动物模型,但这项研究的结果提供了对种间植入物性能差异的深入了解。例如,应该将狗的THR表现与年轻人而不是老年人相比。