The off-loading characteristics of the cast-braces of 30 patients with fractures of the shaft of the femur have been investigated, during axial loading, using strain-gauge transducers. These were applied at the level of the fracture, where the cast was circumferentially split, and to the hinges of the brace at the knee. They measured the load transferred between the two portions of the thigh cast, and between the thigh cast as a whole and the below-knee cast; by subtraction from the total load on the limb, the skeletal force at the fracture level and at the knee could be calculated. In all patients there was an increase in the fracture load as union progressed which was thought to be due to physiological feedback mechanism from the fracture site. The load carried by the two portions of the thigh cast and by the thigh cast as a whole was proportionately high at first and stabilised at an average of 35 per cent of body weight.
机译:在轴向负荷期间,使用应变计换能器研究了30例股骨干骨折患者铸形支架的负重特性。将它们应用在骨折处,在该处骨折在圆周上裂开,并在膝盖处支撑的铰链处使用。他们测量了在大腿模型的两个部分之间以及大腿模型整体与膝盖以下模型之间传递的载荷。通过减去肢体的总负荷,可以计算出骨折水平和膝关节的骨骼力。在所有患者中,随着联合的进行,骨折负荷增加,这被认为是由于来自骨折部位的生理反馈机制所致。大腿石膏的两个部分以及大腿石膏整体所承受的负荷起初是成比例的高,并稳定在平均体重的35%。