A National Outbreak of Ralstonia mannitolilytica Associated With Use of a Contaminated Oxygen-Delivery Device Among Pediatric Patients

机译:在全国范围内,小儿麻痹性罗氏菌的全国爆发与使用受污染的氧气输送装置有关

获取原文

摘要

OBJECTIVES. In August 2005, the Centers for Disease Control and Prevention was notified of a Ralstonia species outbreak among pediatric patients receiving supplemental oxygen therapy with the Vapotherm 2000i (Vapotherm, Inc, Stevensville, MD). The Vapotherm 2000i is a reusable medical device that was used in 900 hospitals in the United States in 2005. Ralstonia are waterborne bacilli that have been implicated in hospital-acquired infections. We initiated an investigation to determine the source of the outbreak and implement infection control and prevention measures.PATIENTS AND METHODS. We performed a case-control study at 1 hospital and conducted national case findings to obtain clinical and environmental samples for laboratory analysis. Case-patients had health care–acquired Ralstonia colonization or infection. Isolates were compared by using pulsed-field gel electrophoresis. We tested manufacturer-recommended disinfection protocols for the Vapotherm 2000i under simulated-use conditions.RESULTS. Case-patients at the hospital ( n = 5) were more likely to have received Vapotherm therapy than controls. Nationally, Ralstonia mannitolilytica was confirmed in 38 patients (aged 5 days to 7 years); 35 (92%) of the patients were exposed to the Vapotherm 2000i before recovery of the organism. Pulsed-field gel electrophoresis showed related R mannitolilytica strains from isolates sent from 18 hospitals in 12 states. A Vapotherm machine reprocessed with a protocol proposed by the manufacturer grew Ralstonia spp after 7 days of simulated use. In December 2005, Vapotherm recalled the 2000i.CONCLUSIONS. Our findings suggest intrinsic contamination of Vapotherm devices with Ralstonia spp. New medical devices may provide therapy equivalent to current devices yet pose novel reprocessing challenges.
机译:目标2005年8月,美国疾病控制与预防中心被告知在接受Vapotherm 2000i辅助供氧治疗的小儿科患者中Ralstonia物种暴发(马里兰州史蒂文斯维尔的Vapotherm公司)。 Vapotherm 2000i是可重复使用的医疗设备,2005年在美国的900多家医院中使用过。Ralstonia是水生细菌,与医院获得性感染有关。我们开始调查以确定爆发的来源并采取感染控制和预防措施。患者和方法。我们在一家医院进行了病例对照研究,并进行了全国性病例调查,以获取临床和环境样本进行实验室分析。案例患者接受过医疗保健,即获得了Ralstonia菌落或感染。通过使用脉冲场凝胶电泳比较分离物。我们在模拟使用条件下测试了制造商推荐的Vapotherm 2000i消毒方案。结果。医院中的病例患者(n = 5)比对照组更容易接受Vapotherm治疗。在全国范围内,已确认38例患者(年龄5天至7岁)中出现了曼尼罗尼亚氏菌;在病原体恢复之前,有35名患者(92%)接触了Vapotherm 2000i。脉冲场凝胶电泳显示了来自12个州的18家医院的分离株中分离的R mannitolilytica菌株。经过7天模拟使用后,经制造商提出的协议重新处理的Vapotherm机器长出了Ralstonia spp。 2005年12月,Vapotherm召回了2000i。我们的研究结果表明,Ralstonia spp对Vapotherm装置造成了内在污染。新的医疗设备可能会提供与当前设备相同的疗法,但却带来了新的后处理挑战。
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号