摘要

A new electrochemical sensor based on a Nafion, aminated reduced graphene oxide and chitosan functional membrane-modified glassy carbon electrode was proposed for the simultaneous detection of adenine and guanine. Fourier transform-infrared spectrometry (FTIR), transmission electron microscopy (TEM), and electrochemical methods were utilized for the additional characterization of the membrane materials. The prepared electrode was utilized for the detection of guanine (G) and adenine (A). The anodic peak currents to G and A were linear in the concentrations ranging from 0.1 to 120 μM and 0.2 to 110 μM, respectively. The detection limits were found to be 0.1 μM and 0.2 μM, respectively. Moreover, the modified electrode could also be used to determine G and A in calf thymus DNA.
机译:提出了一种基于Nafion,胺化还原氧化石墨烯和壳聚糖功能膜修饰的玻碳电极的新型电化学传感器,用于同时检测腺嘌呤和鸟嘌呤。傅里叶变换红外光谱(FTIR),透射电子显微镜(TEM)和电化学方法被用于膜材料的附加表征。所制备的电极用于检测鸟嘌呤(G)和腺嘌呤(A)。到达G和A的阳极峰值电流分别在0.1至120μM和0.2至110μM的浓度范围内呈线性关系。发现检出限分别为0.1μM和0.2μM。此外,修饰的电极还可用于测定小牛胸腺DNA中的G和A。