首页> 外文期刊>Pediatrics: Official Publication of the American Academy of Pediatrics >Electrocardiograph Monitor Artifacts in a Neonatal Intensive Care Unit
【24h】

Electrocardiograph Monitor Artifacts in a Neonatal Intensive Care Unit

机译:新生儿重症监护室中的心电图监护仪伪像

获取原文
           

摘要

A wide variety of artifacts may be found when monitoring the ECG in a neonatal intensive care unit. Many of the artifacts resemble arrhythmias, and unless they are recognized as artifacts they may lead to serious errors of diagnosis and therapy. Many of the artifacts are caused by patient movement such as seizures, tremulousness, or hiccups. Others may be introduced by the monitor itself or be caused by electrical equipment in the vicinity. A group of ECC tracings is presented to illustrate the various artifacts encountered. Features that distinguish the artifacts from the arrhythmias they mimic are described, as are suggestions for elimination of the artifacts.
机译:在新生儿重症监护室中监测ECG时,可能会发现各种各样的伪影。许多伪像类似于心律不齐,并且除非被识别为伪像,否则可能会导致严重的诊断和治疗错误。许多伪像是由患者的运动引起的,例如癫痫发作,颤抖或打h。其他可能是监视器本身引入的,也可能是附近的电气设备引起的。展示了一组ECC跟踪,以说明遇到的各种工件。描述了区分伪影和模仿的心律不齐的功能,以及消除伪影的建议。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有