A study of the frequency response of AFM microcantilevers in liquid media contained in a commercial fluid cell is presented. Such systems exhibit complicated dynamics which are often not well described by available theories. Their dynamic behavior has a direct effect on the use of the AFM in dynamic mode while imaging in liquid or while extracting the rheological properties of the fluid. We explore the issues related to the design of the cantilever holder/fluid cell and propose an approach for evaluating, minimizing and recognizing the ultimate limitations of commercial cantilever holders. A technique for estimating the frequency response spectrum of the fluid cell itself from experimental data is presented. This spectrum can then be used to evaluate whether or not the fluid cell is suited for the desired purpose.
机译:提出了对AFM微悬臂梁在商业流体池中所含液体介质中的频率响应的研究。这样的系统表现出复杂的动力学,而现有理论常常无法很好地描述它们。当在液体中成像或提取流体的流变特性时,它们的动态行为直接影响动态模式下AFM的使用。我们探索与悬臂支架/流体单元的设计有关的问题,并提出一种评估,最小化和认识到商业悬臂支架的最终局限性的方法。提出了一种根据实验数据估算流体池本身的频率响应谱的技术。然后,可以使用该频谱来评估流体池是否适合所需的目的。