首页> 外文期刊>Sensors >How Accurate Are Electronic Monitoring Devices? A Laboratory Study Testing Two Devices to Measure Medication Adherence
【24h】

How Accurate Are Electronic Monitoring Devices? A Laboratory Study Testing Two Devices to Measure Medication Adherence

机译:电子监控设备的准确性如何?一项实验室研究,测试了两种用于测量药物依从性的设备

获取原文
           

摘要

In a prospective descriptive laboratory study, 25 Helping Hand™ (HH) (10 without and 15 with reminder system) and 50 Medication Event Monitoring Systems (MEMS) (25 with 18-month and 25 with 2-year battery life) were manipulated twice daily following a predefined protocol during 3 consecutive weeks. Accuracy was determined using the fixed manipulation scheme as the reference. Perfect functioning (i.e., total absence of missing registrations and/or overregistrations) was observed in 70% of the HH without, 87% of the HH with reminder, 20% MEMS with 18 months, and 100% with 2-year battery life respectively.
机译:在一项前瞻性描述性实验室研究中,使用25个Helping Hand ™(HH)(10个不带提醒系统和15个带提醒系统)和50个药物事件监测系统(MEMS)(25个带18个月的服务和25个带2-电池寿命(一年)(电池续航时间)在连续3周内按照预定方案每天进行两次操作。使用固定操作方案作为参考确定准确性。分别在70%的无HH,87%的HH有提醒,20%的MEMS有18个月和100%的有2年电池寿命中观察到完美的功能(即完全没有丢失的注册和/或过度注册) 。