首页> 外文期刊>Sensors >Effect of Structural Design of Silver/Silver Chloride Electrodes on Stability and Response Time and the Implications for Improved Accuracy in pH Measurement
【24h】

Effect of Structural Design of Silver/Silver Chloride Electrodes on Stability and Response Time and the Implications for Improved Accuracy in pH Measurement

机译:氯化银/氯化银电极的结构设计对稳定性和响应时间的影响以及pH测量精度提高的意义

获取原文
           

摘要

The response time of thermal electrolytic Ag/AgCl reference electrodes is defined by the porous structure that limits the rate at which traces of any previous solution are diluted by any new solution environment. The electrode stabilisation time when transferred from one electrolyte to another has been shown to change when different structural designs to the conventional sphere of Ag/AgCl are used. Electrodes fabricated with cylindrical architectures of Ag/AgCl have shown improved stability and reach equilibrium faster than spherical electrodes. This work has positive implications for the accuracy and throughput of primary pH measurements made using the Harned cell.
机译:热电解Ag / AgCl参比电极的响应时间由多孔结构定义,该结构限制了任何先前溶液的痕迹被任何新溶液环境稀释的速率。当使用与传统的Ag / AgCl球形不同的结构设计时,从一种电解质转移到另一种电解质时的电极稳定时间已显示出变化。与球形电极相比,用圆柱形Ag / AgCl结构制造的电极显示出更高的稳定性并更快达到平衡。这项工作对使用Harned池进行的一次pH测量的准确性和通量具有积极意义。