首页> 外文期刊>Sensors >Laboratory Calibration of a Field Imaging Spectrometer System
【24h】

Laboratory Calibration of a Field Imaging Spectrometer System

机译:现场成像光谱仪系统的实验室校准

获取原文
           

摘要

A new Field Imaging Spectrometer System (FISS) based on a cooling area CCD was developed. This paper describes the imaging principle, structural design, and main parameters of the FISS sensor. The FISS was spectrally calibrated with a double grating monochromator to determine the center wavelength and FWHM of each band. Calibration results showed that the spectral range of the FISS system is 437–902 nm, the number of channels is 344 and the spectral resolution of each channel is better than 5 nm. An integrating sphere was used to achieve absolute radiometric calibration of the FISS with less than 5% calibration error for each band. There are 215 channels with signal to noise ratios (SNRs) greater than 500 (62.5% of the bands). The results demonstrated that the FISS has achieved high performance that assures the feasibility of its practical use in various fields.
机译:开发了基于冷却区域CCD的新型场成像光谱仪系统(FISS)。本文介绍了FISS传感器的成像原理,结构设计和主要参数。使用双光栅单色仪对FISS进行光谱校准,以确定每个波段的中心波长和FWHM。校准结果表明,FISS系统的光谱范围为437–902 nm,通道数为344,每个通道的光谱分辨率均优于5 nm。积分球用于实现FISS的绝对辐射定标,每个波段的校准误差均小于5%。有215个通道的信噪比(SNR)大于500(占频段的62.5%)。结果表明,FISS已获得了高性能,从而确保了其在各个领域中实际应用的可行性。