In this study, we describe a novel solid-state thin-film Ag/AgCl reference electrode (SSRE) that was coated with a protective layer of graphene oxide (GO). This layer was prepared by drop casting a solution of GO on the Ag/AgCl thin film. The potential differences exhibited by the SSRE were less than 2 mV for 26 days. The cyclic voltammograms of the SSRE were almost similar to those of a commercial reference electrode, while the diffusion coefficient of Fe(CN)63− as calculated from the cathodic peaks of the SSRE was 6.48 × 10−6 cm2/s. The SSRE was used in conjunction with a laboratory-made working electrode to determine its suitability for practical use. The average pH sensitivity of this combined sensor was 58.5 mV/pH in the acid-to-base direction; the correlation coefficient was greater than 0.99. In addition, an integrated pH sensor that included the SSRE was packaged in a secure digital (SD) card and tested. The average sensitivity of the chip was 56.8 mV/pH, with the correlation coefficient being greater than 0.99. In addition, a pH sensing test was also performed by using a laboratory-made potentiometer, which showed a sensitivity of 55.4 mV/pH, with the correlation coefficient being greater than 0.99.
机译:在这项研究中,我们描述了一种新型的固态薄膜Ag / AgCl参比电极(SSRE),该参比电极涂有氧化石墨烯(GO)的保护层。通过在Ag / AgCl薄膜上滴铸GO溶液来制备该层。 SSRE表现出的电位差在26天内小于2 mV。 SSRE的循环伏安图几乎与市售参比电极相似,而Fe(CN) 6 3-的扩散系数由SSRE为6.48×10 −6 cm 2 / s。 SSRE与实验室制造的工作电极一起使用,以确定其实用性。该组合传感器在酸碱方向上的平均pH敏感度为58.5 mV / pH。相关系数大于0.99。此外,将包含SSRE的集成pH传感器包装在安全数字(SD)卡中并进行了测试。芯片的平均灵敏度为56.8 mV / pH,相关系数大于0.99。此外,还使用实验室制造的电位计进行了pH感测测试,该电位计的灵敏度为55.4 mV / pH,相关系数大于0.99。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号