Reproducibility of Peripapillary Retinal Nerve Fiber Layer Thickness and Optic Nerve Head Parameters Measured with Cirrus HD-OCT in Glaucomatous Eyes

机译:用Cirrus HD-OCT在青光眼中测量乳头周围视网膜神经纤维层厚度和视神经头参数的再现性

获取原文
           

摘要

Purpose.: To assess the reproducibility of peripapillary retinal nerve fiber layer (RNFL) thickness and optic nerve head (ONH) parameters measured with Cirrus HD-OCT in glaucomatous eyes. Methods.: Fifty-five glaucomatous eyes were included in the study. The optic disc cube 200 ?? 200 protocol was used to obtain three scans during the same visit to evaluate the intravisit reproducibility. One scan on 4 additional days within a 2-month period of the first session was obtained to assess intervisit reproducibility. Intraclass correlation coefficient (ICC), coefficient of variation (CV), and test-retest SD (TRT SD) were calculated for each RNFL and ONH parameter. The formula 1.645 ?? a??2 ?? intervisit TRT SD provides an upper tolerance limit to variability beyond which nonphysiologic change should be considered. Results.: All ICCs were excellent, ranging from 83.9% to 99.2% for intravisit measurements and from 80.8% to 99.1% for intervisit measurements. Cup/disc area ratio had the lowest CV (1.1%) in either type of measurement, followed by average RNFL thickness (1.9% and 2.7%). Nasal clock hours and quadrants showed the poorest reproducibility as did the clock hour directly temporally. The intervisit tolerance limit for average RNFL thickness was 3.89 ??m. Conclusions.: Intravisit and intervisit measurements of peripapillary RNFL thickness and ONH parameters with Cirrus HD-OCT showed excellent reproducibility, indicating that this instrument may be useful in monitoring glaucoma progression. When comparing two measurements from the same eye on two different visits, a reproducible decrease in average RNFL thickness of approximately 4 ??m or more may be considered a statistically significant change from baseline.
机译:目的:评估用Cirrus HD-OCT在青光眼中测量的乳头周围视网膜神经纤维层(RNFL)厚度和视神经头(ONH)参数的可再现性。方法:研究纳入了55只青光眼。视盘立方体200 ??在同一次就诊期间,使用200协议获得三次扫描,以评估内部访问的可重复性。在第一届会议的两个月内,对另外4天进行了一次扫描,以评估间次重复性。针对每个RNFL和ONH参数计算了类内相关系数(ICC),变异系数(CV)和重测SD(TRT SD)。公式1.645 ?? a2 ?? Intervisit TRT SD为变异性提供了一个上限公差,超出该上限应考虑非生理变化。结果:所有ICC都非常出色,对于内视测量而言,范围从83.9%到99.2%,对于间视测量范围从80.8%到99.1%。在两种类型的测量中,杯/盘面积比的CV最低(1.1%),其次是RNFL的平均厚度(1.9%和2.7%)。鼻时钟时间和象限显示的可重复性最差,而时钟时间直接在时间上也是如此。平均RNFL厚度的间隙公差极限为3.89 ?? m。结论:使用Cirrus HD-OCT进行的视乳头周围和鼻间RNFL厚度和ONH参数的内视和间视测量显示出极好的可重复性,表明该仪器可用于监测青光眼的进展。当在两次不同的访问中比较同一只眼睛的两次测量结果时,平均RNFL厚度可再现性降低约4?m或更多可被视为与基线相比有统计学意义的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号