Feasibility Study of the Use of Frozen Human Sera in Split-Sample Comparison of Immunoassays with Candidate Reference Measurement Procedures for Total Thyroxine and Total Triiodothyronine Measurements

机译:使用冷冻人血清进行免疫测定的分样品比较的可行性研究,并采用候选参考测量程序进行总甲状腺素和总三碘甲状腺素的测量

获取原文
           

摘要

Background: Diagnostic manufacturers must ensure/document metrologically traceable assays. We report on a feasibility study of a split-sample comparison for that purpose. Processed, frozen single-donation sera, assigned target values by candidate reference measurement procedures (cRMPs), were used with immunoassays for total thyroxine (TT4) and triiodothyronine (TT3) as models.Methods: Two serum panels were quantified for TT4 and TT3 with validated cRMPs and measured in parallel with at least 14 immunoassays. The results were interpreted in terms of traceability of calibration (trueness) and of the individual measurement result (accuracy) by linear regression analysis and graphical representation against specifications. The commutability of the sera was investigated by parallel analysis of TT4 in freshly collected but nonfiltered specimens.Results: The TT4 (TT3) concentrations in the sera (according to the cRMPs) were 64–269 nmol/L (0.88–13.7 nmol/L). The method comparison showed that for TT4, on average, the immunoassays produced results in agreement with the cRMPs, whereas for TT3, results were typically higher. It also demonstrated a considerable between-assay divergence in traceability of calibration and accuracy. The evidence of noncommutability of the sera attributable to processing, however, indicates that the interpretation should be treated with caution.Conclusions: Frozen sera can be used for documenting/validating traceability of total thyroid measurements. The way in which the sera are processed may jeopardize commutability, however, and therefore requires in-depth investigation.
机译:背景:诊断制造商必须确保/记录可溯源的测定。我们报告了为此目的进行拆分样本比较的可行性研究。处理过的冷冻单次献血血清,通过候选参考测量程序(cRMPs)分配了目标值,并采用免疫测定法作为模型中的总甲状腺素(TT4)和三碘甲状腺素(TT3)。经验证的cRMP,并与至少14次免疫测定平行进行。通过线性回归分析和针对规格的图形表示,以校准的可追溯性(真实性)和单个测量结果的准确性(准确性)来解释结果。通过对新鲜采集但未过滤的标本中的TT4进行平行分析,研究了血清的可交换性。结果:血清中TT4(TT3)浓度(根据cRMPs)为64–269 nmol / L(0.88–13.7 nmol / L) )。方法比较表明,对于TT4,平均而言,免疫分析产生的结果与cRMP一致,而对于TT3,则通常更高。它也显示出校准和准确性的可追溯性之间存在显着的测定间差异。归因于加工过程的血清不可交换性的证据表明,应谨慎对待解释。结论:冷冻血清可用于记录/验证总甲状腺测量结果的可追溯性。但是,处理血清的方式可能会危害可交换性,因此需要进行深入研究。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号