Through autoverification, a customized expert system within the Laboratory Information System (LIS), a computer performs the initial review and verification of test results based on a predetermined set of boundaries or rules, as established by the laboratory (1). A carefully designed system can be an important tool in addressing such crucial issues as medical errors, test turnaround time (TAT), shortages in personnel and funding, and operational efficiency. As part of an ongoing modular laboratory automation plan at John H. Stroger, Jr. Hospital of Cook County (JHSHCC, formerly Cook County Hospital, Chicago), autoverification was implemented in our clinical chemistry and urinalysis laboratories.JHSHCC is a large, urban, tertiary acute-care public hospital and trauma center that serves as the centerpiece of a countywide integrated healthcare delivery system. As part of the hospital’s clinical pathology service, the clinical chemistry laboratory provides more than 150 different test procedures in general chemistry, toxicology, drug screening, endocrinology, urinalysis, and blood gas analysis. The laboratory has modern, automated instrumentation, including preanalytical specimen processors (TECAN FE500), all of which are bidirectionally interfaced to the LIS (Soft Computer Consultants). In 2003, the clinical chemistry laboratory processed 1.3 million specimens (more than 6 million reportable results). The result-review/reporting process for the above workload required 14 full-time equivalents (FTE) of technical personnel. Autoverification was expected to reduce the FTEs necessary for this process.Autoverification setup and implementation required 2 FTEs in personnel over a period of 6 weeks. Automated chemistry assays were chosen for the initial autoverification attempt for several reasons, including high test volume, clearly defined reference ranges and delta checks, ability to define intermediate ranges for abnormal results, and the flagging capabilities of various automated analyzers. Urinalysis was included for the autoverification setup based on the availability of interfaced automated analyzers (IRIS 939 Udx) with which decision rules could …
机译:通过自动验证(实验室信息系统(LIS)中的定制专家系统),计算机可以根据实验室(1)建立的一组预定边界或规则对测试结果进行初始检查和验证。经过精心设计的系统可能是解决诸如医疗错误,测试周转时间(TAT),人员和资金短缺以及运营效率等关键问题的重要工具。作为库克县小医院John H.Stroger(JHSHCC,前身是芝加哥库克县医院)正在进行的模块化实验室自动化计划的一部分,在我们的临床化学和尿液实验室中实施了自动验证。三级急诊公立医院和创伤中心,是全县范围内综合医疗服务系统的核心。作为医院临床病理服务的一部分,临床化学实验室在常规化学,毒理学,药物筛选,内分泌学,尿液分析和血气分析方面提供了150多种不同的测试程序。实验室拥有现代化的自动化仪器,包括分析前样品处理机(TECAN FE500),所有这些仪器都与LIS(软计算机顾问)双向接口。 2003年,临床化学实验室处理了130万个标本(报告的结果超过600万个)。上述工作量的结果审核/报告过程需要14个等价的全职技术人员(FTE)。预计自动验证将减少此过程所需的FTE。自动验证的设置和实施需要在6周内为人员配备2个FTE。选择自动化学分析进行初始自动验证的尝试有几个原因,包括高测试量,明确定义的参考范围和变化量检查,为异常结果定义中间范围的能力以及各种自动化分析仪的标记功能。基于接口自动分析仪(IRIS 939 Udx)的可用性,可将尿液分析包括在自动验证设置中,通过该分析器,决策规则可以…

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号