首页> 外文期刊>Clinical Chemistry: Journal of the American Association for Clinical Chemists >Precipitation methods for high-density lipoprotein cholesterol measurement compared, and final evaluation under routine operating conditions of a method with a low sample-to-reagent ratio
【24h】

Precipitation methods for high-density lipoprotein cholesterol measurement compared, and final evaluation under routine operating conditions of a method with a low sample-to-reagent ratio

机译:比较了用于高密度脂蛋白胆固醇测量的沉淀方法,以及在常规操作条件下对样品/试剂比率低的方法进行的最终评估

获取原文
           

摘要

We evaluated six precipitation methods for high-density lipoprotein cholesterol (HDL-chol) determination: the heparin/Mn2+precipitation reagent method (Hep), two variants of the phosphotungstic acid/Mg2+ method (Tung-L and Tung-B), the dextran sulfate 50 000/Mg2+ method (Dex), the PEG 6000 method (PEG), and the PEG 6000/dextran sulfate 15 000 (PEG/Dex) method. The Tung-B and PEG/Dex precipitation methods have a low sample/precipitation reagent volume ratio (0.4). The Tung-B, Dex, PEG, and PEG/Dex methods gave similar values, averaging within 0.1 mmol/L of each other, showing that the precipitation selectivity of these methods is comparable. The precipitation efficiency of Tung-B and Peg/Dex, however, was superior. Ultrafiltration of the supernatants was needed only at triglyceride concentrations 16.4 mmol/L (undiluted sample) or 28.0 mmol/L (sample diluted twofold); however, ultrafiltration without dilution was the most accurate method. Results of Tung-B under routine conditions (33 technicians) agreed well with those of the PEG method for 406 normo- and hyperlipidemic plasma samples. By comparison with the HDL-chol method from the Centers for Disease Control and Prevention, the Tung-B method showed a total error of 10.6%, which fulfills the criteria of the National Cholesterol Education Program for HDL-chol analysis. In conclusion, with motivated personnel, Tung-B is a reliable, cost-effective method for routine HDL-chol analysis.
机译:我们评估了六种沉淀法用于高密度脂蛋白胆固醇(HDL-chol)的测定:肝素/ Mn2 +沉淀试剂法(Hep),磷酸钨酸/ Mg2 +方法的两种变体(Tung-L和Tung-B),硫酸葡聚糖50000 / Mg2 +方法(Dex),PEG 6000方法(PEG)和PEG 6000 /硫酸葡聚糖15000(PEG / Dex)方法。 Tung-B和PEG / Dex沉淀方法的样品/沉淀试剂体积比低(<0.4)。 Tung-B,Dex,PEG和PEG / Dex方法给出的值相似,平均值平均在0.1 mmol / L之内,表明这些方法的沉淀选择性相当。然而,Tung-B和Peg / Dex的沉淀效率更高。仅在甘油三酯浓度> 16.4 mmol / L(未稀释样品)或> 28.0 mmol / L(样品稀释两倍)时才需要对上清液进行超滤。然而,不稀释的超滤是最准确的方法。在常规条件下,Tung-B的结果(33名技术人员)与PEG方法对406个正常和高血脂血浆样品的结果非常吻合。通过与疾病控制和预防中心的HDL-chol方法进行比较,Tung-B方法显示总误差为10.6%,符合国家胆固醇教育计划对HDL-chol分析的标准。总而言之,在积极进取的工作人员的帮助下,Tung-B是一种可靠的,具有成本效益的方法,用于常规HDL-chol分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号