Compact analyzers suited to near-patient testing estimate hematocrit by measuring the conductivity of undiluted blood. We evaluated the accuracy of hematocrit determination of one such analyzer (Instrumentation Laboratory BGE analyzer) against an automated cell counter (EPC) and packed cell volume (PCV) microhematocrit. When specimens (n = 34) from outpatient and ward patients were analyzed with all three methods, the BGE analyzer correlated well with both EPC and PCV hematocrit determinations (BGE = 1.00 PCV + 0.3%, S(y)/x = 2.0%), suggesting that all three methods are similar in performance for most patients. However, a patient with increased plasma osmolality showed significant decreases in BGE and PCV hematocrits relative to the EPC method. The differences in hematocrit measurements could be reproduced by adding solutes to blood in vitro or by modifying the plasma osmolality of rats in vivo. Samples from patients undergoing cardiac surgery, whose blood had large changes in protein concentration, showed discrepancies between hematocrits by conductivity and other methods; similar effects could be produced by changes in protein concentration or in vitro addition of polyethylene glycol. We conclude that conductivity measurements provide accurate hematocrit results for physiologically normal subjects but not for some intensive-care and surgical patients.
机译:紧凑型分析仪可通过测量未稀释血液的电导率来估算血细胞比容,适用于近乎患者的测试。我们评估了一种此类分析仪(仪器实验室BGE分析仪)相对于自动细胞计数器(EPC)和包装细胞体积(PCV)微量血细胞比容的血细胞比容测定的准确性。使用三种方法对门诊和病房患者的标本(n = 34)进行分析时,BGE分析仪与EPC和PCV血细胞比容测定值之间的相关性很好(BGE = 1.00 PCV + 0.3%,S(y)/ x = 2.0%) ,这表明这三种方法对大多数患者的表现均相似。但是,相对于EPC方法,血浆渗透压升高的患者BGE和PCV血细胞比容显着降低。血细胞比容测量值的差异可以通过在体外向血液中添加溶质或在体内改变大鼠的血浆渗透压来重现。心脏手术患者的血液中蛋白质浓度发生了很大的变化,通过电导率和其他方法显示血细胞比容之间存在差异。通过改变蛋白质浓度或在体外添加聚乙二醇可以产生类似的效果。我们得出的结论是,电导率测量可为生理正常受试者提供准确的血细胞比容结果,但不能为某些重症监护和手术患者提供准确的血细胞比容结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号